45
Doc. No.
:
P30001PE0101-07
9.4.Test Screen
Open the Test Screen by pressing the
“3” key from the MENU Screen.
Press the
“ESC” key from the Test Screen to return to the MENU Screen.
9.4.1.
Axis Test Screen
Open the Axis Test Screen by pressing the
“1” key from the Test Screen.
・
“Start” key
:
Press this key to check the operation of all axes. Use this function to
determine where a problem occurred during a run. Place all required disposables onto
the instrument before executing this test.
・
“ESC” key
:
Press this key to display the Test Screen.
9.4.2.
Temp Test Screen
Open the Temp Test
Screen by pressing “2” key from the Test Screen.
“SS.S” indicates the set temperature in degrees Celsius. Press
the up arrow or the down arrow to increase or decrease the
temperature. The upper limit is 99°C. Press “START” to start
the heating process.
The following screen appears:
DD MM YYYY HH:MM
START:Run 1:UV
2:Man 3:Test 4:Setup
Key:START,1,2,3,4
TEST
1:Axis 2:Temp
3:Serial 4:Version
Key:1,2,3,4,ESC
TEST:AXIS
insert disposables
START:Run
Key:1,2,3,4,ESC
TEST: TEMPERATURE
set temp: SS.S C
START: Run
Key:Up,Dn,START,ESC