7 - EN
1. BP MANAGER software CD-ROM
2. connecting cable (USB)
3. instruction manual
are packaged and sold separately from the blood pressure monitor.
Troubleshooting
If any abnormality should arise during use, please check the following points.
EE / Measurement Error:
Exceeding measurement range. Measure again. Wrap the cuff cor-
rectly and keep wrist steady during measurement.
E1 / Air Circuit Abnormality:
Measure again.
E2 / Pressure Exceeding 300 mmHg:
Switch the unit off clear, then measure again.
E3 / Data Error:
Remove and reload the batteries.
Er / Exceeding Measurement Range:
Measure again.
No display when the START/STOP key is pressed:
Re-insert the batteries in the correct posi-
tions.
Note:
If the unit still does not work, return it to your dealer. Under no circumstance should you
disassemble and repair the unit by yourself.
Cautionary Notes
1. The unit contains high-precision assemblies. Therefore, avoid extreme temperatures, humidity,
and direct sunlight. Avoid dropping or strongly shocking the main unit, and protect it from
dust.
2. Clean the blood pressure monitor body and the cuff carefully with a slightly damp, soft cloth.
Do not press. Do not wash the cuff or use chemical cleaner on it. Never use thinner, alcohol or
petrol (gasoline) as cleaner.
3. Leaky batteries can damage the unit. Remove the batteries when the unit is not used for a
long time.
4. The unit should not be operated by children so to avoid hazardous situations.
5. If the unit is stored near freezing, allow it to acclimate at room temperature before use.
6. This unit is not field serviceable. You should not use any tool to open the device nor should you
attempt to adjust anything inside the device. If you have any problems, please contact the store
or the doctor from whom you purchased this unit or please contact Rossmax International Ltd.
7. For users diagnosed with common arrhythmia (atrial or ventricular premature beats or atrial
fibrillation), diabetes, poor circulation of blood, kidney problems, or for users suffered from
stroke, or for unconscious users, the device may have difficulty in determining the proper blood
pressure.
8. To stop operation at any time, press the ON/OFF/START key, and the air in the cuff will be rap-
idly exhausted.
9. Once the inflation reaches 300 mmHg, the unit will start deflating rapidly for safety reasons.
10. Please note that this is a home healthcare product only and it is not intended to serve as a