Pressure Limits
See Table A-6, Table A-7, and Table A-8.
Vacuum Limits
PTFE Lining
Full vacuum to 350 °F (177 °C) through 4-in. (100 mm) line sizes. Consult
factory for vacuum applications with line sizes of 6 inches (150 mm) or
larger.
All Other Standard Sensor Lining Materials
Full vacuum to maximum material temperature limits for all available line
sizes.
Submergence Protection
(1)
IP68. Continuous submergence to 30 ft. (10 m) requires that conduit entries of
the sensor remote junction box be properly sealed to prevent water ingress.
This requires the user to install sealed IP68 approved cable glands, conduit
connections, or conduit plugs. For more details on proper installation
techniques for an IP68 / submersible application, reference Rosemount
Technical Document 00840-0100-4750 available on www.Rosemount.com.
Conductivity Limits
Process liquid must have a conductivity of 5 microsiemens/cm (5
micromhos/cm) or greater for 8705. Process liquid must have a conductivity of
50 microsiemens/cm (50 micromhos/cm) for 8707 when used with 8712H, 5
microsiemens/cm when used with other transmitters. Excludes the effect of
interconnecting cable length in remote mount transmitter installations.
(1)
Available for Rosemount 8705 Sensors only.
Table A-6. Temperature vs. Pressure Limits
(1)
Sensor Temperature vs. Pressure Limits for ASME B16.5 Class Flanges (
1
/
2
- to 36-in. line sizes)
(2)
Pressure
Flange Material
Flange Rating
@ -20 to 100
°
F
(-29 to 38
°
C)
@ 200
°
F
(93
°
C)
@ 300
°
F
(149
°
C)
@ 350
°
F
(177
°
C)
Carbon Steel
Class 150
285 psi
260 psi
230 psi
215 psi
Class 300
740 psi
675 psi
655 psi
645 psi
Class 600
(3)
1000 psi
800 psi
700 psi
650 psi
Class 600
(4)
1480 psi
1350 psi
1315 psi
1292 psi
Class 900
2220 psi
2025 psi
1970 psi
1935 psi
Class 1500
3705 psi
3375 psi
3280 psi
3225 psi
Class 2500
6170 psi
5625 psi
5470 psi
5375 psi
304 Stainless Steel
Class 150
275 psi
235 psi
205 psi
190 psi
Class 300
720 psi
600 psi
530 psi
500 psi
Class 600
(5)
1000 psi
800 psi
700 psi
650 psi
Class 600
(6)
1440 psi
1200 psi
1055 psi
997 psi
Class 900
2160 psi
1800 psi
1585 psi
1497 psi
Class 1500
3600 psi
3000 psi
2640 psi
2495 psi
Class 2500
6000 psi
5000 psi
4400 psi
4160 psi
(1) Liner temperature limits must also be considered. Polyurethane, Linatex, and Neoprene have temperature limits of 140 °F (60 °C), 158 °F (70 °C), and 176