Quick Start Guide
25
September 2016
Special Conditions for Safe Use (X):
1. If the equipment is fitted with an optional 90V transient suppressor, it is not capable of
withstanding the 500V insulation test required by ABNT NBR IRC 60079-11. This must
be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given a protective polyurethane
paint finish; however, care should be taken to protect it from impact or abrasion i
n
zones that require EPL Ga.
Technical Regulations Customs Union (EAC)
EM
EAC Flameproof
Certificate: RU C-US.
M
ю
62.B.02349
Markings: Ga/Gb Ex d IIC T6…T4 X, T5/T4(-50 °C
≤
T
a
≤
+80 °C), T6(-50 °C
≤
T
a
≤
+65 °C)
Special Condition for Safe Use (X):
See certificate for special conditions
IM
EAC Intrinsically Safe
Certificate: RU C-US.
M
ю
62.B.02349
Markings: 0Ex ia IIB T4 Ga X, T4(-60 °C
≤
T
a
≤
+70 °C)
Special Condition for Safe Use (X):
See certificate for special conditions
Combinations
K1
Combination of E1, I1, N1, and ND
K2
Combination of E2 and I2
K5
Combination of E5 and I5
K6
Combination of E6 and I6
K7
Combination of E7, I7, N7, and NK
KA
Combination of E1, I1, E6, and I6
KB
Combination of E5, I5, E6, and I6
KC
Combination of E1, I1, E5, and I5
KD
Combination of E1, I1, E5, I5, E6, and I6
KM
Combination of EM and IM
Supply
Modbus
RTD
Voltage U
i
22 V
9 V
15.51 V
Current I
i
147 mA
26 mA
20.89 mA
Power P
i
1 W
1 W
80.94 mW
Capacitance C
i
0
0
0
Inductance L
i
0
0
0
Содержание 4088A MultiVariable
Страница 26: ...September 2016 26 Quick Start Guide Figure 16 Rosemount 4088 Declaration of Conformity ...
Страница 27: ...Quick Start Guide 27 September 2016 ...
Страница 28: ...September 2016 28 Quick Start Guide ...
Страница 29: ...Quick Start Guide 29 September 2016 ...
Страница 31: ...Quick Start Guide 31 September 2016 ...