Page 18
Electrical components
Wiring circuit
The nursing table comes with control box mounted on the frame. The diagram
shows how to connect the cables with the control box.
All plug-in connections are waterproof, and mounting/dismounting may therefore
be a little difficult. Make sure that all plugs have the correct position before
inserting them into the control unit. The plugs must be pressed home until the O-
ring is no longer visible.
Motor and control box has an IPX6 rating. This means that they are dust-and
waterproof.
The electrical system is designed for intermittent use, which means a maximum duty
cycle of 10%; 2min on / 18min from.
1.
Controlbox
2.
Mains cable 230V
3.
Actuator, connects to 1
4.
Handset
5.
Safety stop
1
5
2
3
4