Mounting Instruction 

5. Connection of the safety stop 

 

Содержание Maxi 2

Страница 1: ...Ropox Nursing table Maxi 2 Usermanual Mounting Instruction This manual need to follow the product PDF 6140 28 05 2018 ...

Страница 2: ...ctions 7 1 Mounting to the wall 8 2 Mounting of the lower cover 11 3 Mounting of the upper cover 11 4 Connection of the worktop 12 5 Connection of the safety stop 13 6 Connection of water and drain 14 Wiring circuit 18 Connections 19 Dismounting 19 Performance test 20 Labelling 21 Label 21 Symbols 21 Support 22 Troubleshooting 22 Waste 22 Factors affecting the product during storage and transport ...

Страница 3: ...g children up to 75 kg The nursing table can be used I private homes daycare centers or similar institutions to ensure correct and optimal working conditions for staff The height adjustment is performed by pressing the up and down buttons on the handset Item number Lenght of the nursing table Part no 90cm 50 50672 130cm 50 50670 It is important to read this manual before mounting and use of the nu...

Страница 4: ...ctives It has been tested according to European standards DS EN ISO 12182 and DS EN ISO the 60335th Daily use Before using this product it is important that the main power is applied as described under electrical connections The height of the nursing table is adjusted by pressing the up and down buttons on the handset Max load on the nursing table 75kg Warning During height adjustment of the nursi...

Страница 5: ...ake sure that no children are playing under or with the table Make sure that children cannot operate the nursing table unintentionally See that there is free space above and below the table to allow height adjustment Make sure that there are no obstacles above below or around the nursing table in order to avoid the risk of getting jammed during height adjustment If the handset is damaged or does n...

Страница 6: ...s of the nursing table as shown below Placement of 230V socket hot and cold water and drain If possible socket for 230V and drain and connection for hot and cold water should be located as shown on the above sketch The length of the mains cable is 3 2 m Nursing table Maxi 2 90 cm Nursing table maxi 2 130 cm 70 cm ...

Страница 7: ...Dimension sketch These mounting instructions comprise Lifting unit x1 Mounting screws x7 Covers for the lifting unit x1 Rawlplugs x7 Worktop x1 Bolts M5 x 25 x6 Bolt M4 x 10 x9 Safety plate x1 Bolts M8 x 20 x2 Drain and water hose kit x1 Cable Safety stop x1 ...

Страница 8: ...es in the top and four in the bottom of the frame Place the frame against the wall and make sure it is horizontal The frame is standing on two adjustable feet The holes in the frame are all Ø 10 mm Remove the frame and drill the holes Force for each of the four screws in top of the unit Fscrew 75 kg 830 790 4 screws Fscrew 20 kg for each screw If you have other objects mounted right next to this n...

Страница 9: ...Page 9 Mounting instruction Mounting to the wall ...

Страница 10: ...Page 10 Mounting instruction ...

Страница 11: ...Page 11 Mounting instruction 2 Mounting of the lower cover 3 Mounting of the upper cover ...

Страница 12: ...Page 12 Mounting instruction 4 Connection of the worktop 5 Mounting of the worktop ...

Страница 13: ...Page 13 Mounting Instruction 5 Connection of the safety stop ...

Страница 14: ...Page 14 Mounting Instruction 6 Connection of water and drain ...

Страница 15: ...Page 15 Mounting Instruction ...

Страница 16: ...Page 16 Mounting Instruction 7 If the table top is not in horizontal level ...

Страница 17: ...Page 17 ...

Страница 18: ...difficult Make sure that all plugs have the correct position before inserting them into the control unit The plugs must be pressed home until the O ring is no longer visible Motor and control box has an IPX6 rating This means that they are dust and waterproof The electrical system is designed for intermittent use which means a maximum duty cycle of 10 2min on 18min from 1 Controlbox 2 Mains cable ...

Страница 19: ...1 Ports 2 3 4 are not used and are therefore provided by blind plug Handset must be connected to HB The mains cable must be connected to the port on the side of the control box Dismounting Just pull out the mains cable In order to maintain the full IPX6 protection class it is very important that all cables have been mounted correctly ...

Страница 20: ...ove frame to bottom position check movement is even and controlled Be aware that hoes connections follow movement of frame and do not get trapped 2 Press UPWARDS button move frame to top position check movement is even and controlled Also be aware that tabletop does not bump into wall Should this be the case it could be because the wall is skewed If the wall is only a little skewed 1 2mm it is pos...

Страница 21: ...oduct meets the requirements for General Safety and Functionality of the current EU Directives and Standards EMC information The nursing table meets the requirements for Electromagnetic Compatibility EMC if installation and daily use are in accordance with the instructions in this manual On rare occasions the nursing table can be affected by electronic signals from e g laptops mobile phones or the...

Страница 22: ...opox for support Waste The used equipment must be disposed of in accordance with national law We recommend disassembling of the product to the greatest possible extent so that most parts may be recycled Please note that this product contains electrical components Examples of sorting groups Metal plastic cables electronics and parts for destruction The nursing table may be used at temperatures from...

Страница 23: ...the end of this manual The nursing table is maintenance free For safety reasons we recommend regular inspection of all bolts and moving parts to ensure correct fastening to the wall Cleaning Do not use a high pressure washer to clean this product Simply use a soft cloth or sponge with soapy water of max 60 C pH 7 or cleaning alcohol Always clean the nursing table after use to prevent dirt from sti...

Страница 24: ...Page 24 Spare parts For a complete parts list please contact Ropox 97000462 Actuator Type LA31 3 Stroke 300 mm 97000469 Controlbox Type CB9096 2 97000247 Control switch ...

Страница 25: ...ng Table 130cm grey 88 x 75 x 4 cm 50 50685 Adaptor frame for Nursing Table Maxi 2 Wall distance 20 mm 50 50686 Adaptor frame for Nursing Table Maxi 2 Wall distance 80 mm 50 50682 Side covering white compact laminate 82 x 105 cm For mounting on floor wall 50 50683 Side covering white compact laminate 61 x 105 cm For mounting on floor wall ...

Страница 26: ...ed Guard for Medi2 Maxi2 90 cm 82 x 24 cm Mounted on worktop Not for retrofitting On picture there are mounted 2 pcs 50 14684 Mixer tab Non touch monobloc tap Iqua 50 50694 Mixer tab with extension shower head 50 50702 2 step ladder White 50 50690 Extra hand control with spiral cable For Maxi 2 og Medi 2 with safety stop ...

Страница 27: ...Page 27 50 50697 Ropox Basket unit without wash basin 50 50696 Ropox Basket unit with wash basin ...

Страница 28: ...and comply with the following directives and standards DIRECTIVES 2006 42 EF Machinery Directive concerning Machines amended by Directive No 98 37 EG 2014 35 EU Low Voltage directive amended by directive No 2006 95 EU 2014 30 EU EMC directive amended by directive No 2004 108 EU STANDARDS DS EN 1041 2009 DS EN ISO 60335 1 2012 DS EN 14971 2008 DS EN 12221 1 2008 A1 2013 DS EN 12182 2012 DS EN 71 3 ...

Страница 29: ...ved Tel 45 55 75 05 00 Fax 45 55 78 05 50 E mail info ropox dk www ropox de Ropox A S Ringstedgade 221 DK 4700 Naestved Tel 45 55 75 05 00 Fax 45 55 78 05 50 E mail info ropox dk www ropox com Ropox A S Ringstedgade 221 DK 4700 Næstved Tel 45 55 75 05 00 Fax 45 55 78 05 50 E mail info ropox dk www ropox dk ...