Night Setback Input
The Rooster Sensor100 allows the user to connect to the digital input (0V, dry contact), on Input Connector,
positions 5 & 6. The purpose of the Night Setback input is to allow the user the ability to send a remote signal
to all Rooster Sensor100
’s, simultaneously, which chan
ges their behavior during low flow conditions.
This signal is normally initiated by a Laboratory or Building Management System (BMS).
Once the wiring connections are made, the User can assert and remove the night Setback signal from the BMS,
and three Sensor100 states can be selected.
o
On:
This state can be enabled to trigger NSB mode of operation when no NSB trip threshold is
configured.
o
Audible:
In this state, when a night setback signal is asserted, the Sensor100 will still audibly alarm
when airflow drops below the NSB trip threshold.
o
Muted:
In this state, when the night Setback signal is asserted, the Sensor100 will alarm when airflow
drops below the NSB trip threshold, but the alarm will be muted.
o
Off (default):
In this state, when the night Setback signal is asserted, the Sensor100 will ignore it.
Boot Up Procedure
The Rooster
TM
does not have a power-on switch and will become energized as soon as
you plug it in. The Rooster
TM
boots to the screen depicted on the right, with a slot for
a custom logo. The boot screen is designed to support a custom image or logo file in
portable network graphic (.png) format with (240x320) dimensions. This feature must
be requested and is not setup for manual configuration.
Once the start procedure has completed, the Home screen will appear with two
system buttons and you will need to setup password-protected access tiers for
advanced system functions.
This is explained in the “Access Tier Privileges” section
below.
You can reboot your Rooster
TM
Sensor100 at any time by pushing the hard
“
Home
”
and
“
Mute
”
Buttons
simultaneously until the Boot Screen reappears.
Access Tier Privileges: USER, EH&S & CERTIFIER
Users have the ability to customize their containment cabinets within
a defined set of options for each authorization level. There are three
levels of access to the Rooster
TM
: User, EH&S/Facility Manager and
Certifier. These are managed by logging in with a 4-digit numerical
passcode. The User access level does not require a passcode.
Access Privileges Defined
Function
Passcode Required
Adjust Alarm Volume
None
Alarm Latching Settings
EH&S, Certifier
Alarm Ringback Settings
EH&S, Certifier
Mute Live Alarm
None
Set Asset Tag
EH&S
Manufacturer default passcodes:
EH&S
=
8377
Certifier
=
6425