Startup Delay 

The startup delay prevents an alarm from triggering until the workstation and ventilation systems have fully 
powered up and are operational. Refer to chart on next page for further definitions. 
 
Alarm Delay  

Definition 

30s 

Alarm is paused for 30 seconds on power up. 

1 min 

Alarm is paused for 1 minute on power up.

 

2 min 

Alarm is paused for 2 minutes on power up.

 

5 min 

Alarm is paused for 5 minutes on power up.

 

 

Duct Size 

Volumetric flow can be displayed on the home screen for the 
Rooster

TM

 Sensor100. In order to set this up, enter the duct 

diameter screen from the settings menu. Select the shape of 
the duct, and the diameter (in cm). 
 
Once the duct diameter is entered, the user is then able to 
cycle between fpm, m/s, and cfm on the home screen. 

Home>Settings>Duct Size 

Airflow Resolution 

For advanced users, displayed velocity units can be configured to display in an alternate resolution.   
The default resolution for units is 1 fpm and .01 m/s respectively, however, for users who prefer a more  
uniform readout, this can be configured up to 5 fpm or .05 m/s. 

Home>Settings>System>Airflow Settings>Airflow Resolution 
 

Airflow Resolution   Definition 

1 fpm 

Displayed units in FPM are rounded to 1 fpm. This is the default setting. 

2 fpm 

Displayed units in FPM are rounded to 2 fpm. 

5 fpm 

Displayed units in FPM are rounded to 5 fpm. 

.01 m/s 

Displayed units in m/s are rounded to .01 m/s. This is the default setting. 

.02 m/s 

Displayed units in m/s are rounded to .02 m/s. 

.05 m/s 

Displayed units in m/s are rounded to .05 m/s. 

 

Airflow Sensitivity 

This setting smooths airflow readings by applying a filter that determines “sensitivity” of readings. 

 

Home>Settings>System>Airflow Settings>Sensitivity 
 

Sensitivity 

Definition 

Low 

Decreases filtration effect, resulting in greater fluctuation of readings.  

Med 

Default setting. Filtration effect stabilizes readings. 

High 

Increases filtration effect, providing smoother readings. 

       

       

Содержание Sensor 100

Страница 1: ...proprietary information of Degree Controls Inc contents are protected under US copyright laws Degree Controls Inc 2016 High Performance Air Velocity and Temperature Sensor with Full Color Touch Screen...

Страница 2: ...RoosterTM Sensor100 8 Probe Sensor 8 Night Setback Input 8 Boot Up Procedure 9 Access Tier Privileges USER EH S CERTIFIER 9 Set Change Passwords 10 Alarm Threshold Configuration 10 Alarm Troubleshoot...

Страница 3: ...it display and contains many simple to use features and configuration capabilities built into an intuitive user interface that requires little to no training to learn Password protected access tiers e...

Страница 4: ...through low medium or high alarm volumes If logged in as an EH S or Certifier a global minimum volume threshold can be set This restricts a standard User from setting an alarm volume too low for parti...

Страница 5: ...ust be closed See Page 10 for instructions on how to setup night setback alarm thresholds Home Settings System Alarms Sound Alarm Controls Night Setback Alarm Latching EH S or Certifier users can setu...

Страница 6: ...Operating temperature 40 F 140 F 5 C 60 C Storage temperature 40 F 185 F 40 C 85 C Weight 6oz 170g Compliance Standards CE RoHS Rear Panel View of RoosterTM Sensor100 Connection Description Connector...

Страница 7: ...er to be recessed and sit approximately 0 6 15mm proud of the mounting surface 1 Wall mount backplate For installations where the Rooster Sensor100 is mounted directly to a flat surface the Wall mount...

Страница 8: ...as pictured above The RoosterTM Sensor100 has additional input and output capabilities for night setback and alarm output relay The IO connectors present on the RoosterTM Sensor100 have some pins deac...

Страница 9: ...The RoosterTM does not have a power on switch and will become energized as soon as you plug it in The RoosterTM boots to the screen depicted on the right with a slot for a custom logo The boot screen...

Страница 10: ...ust Thresholds 2 To set alarm thresholds you will first have to define an alarm trip point and an alarm clear point The alarm trip point is the air velocity value that will trigger an alarm The alarm...

Страница 11: ...rflow drops below NSB trip point To adjust Night Setback alarm thresholds you must go to the Threshold menu and select Adjust Thresholds Muted Enters Night Setback mode whenever relay is closed and tr...

Страница 12: ...ccurred If the latch alarm setting is set to Password mode an EH S passcode must be entered to unlatch this alarm If the Latch alarm setting is set to ON simply tap the banner to disable the alarm Swi...

Страница 13: ...anced users displayed velocity units can be configured to display in an alternate resolution The default resolution for units is 1 fpm and 01 m s respectively however for users who prefer a more unifo...

Страница 14: ...urther information Home Settings System Advanced I O Polarity DAC Output The RoosterTM Sensor100 has DAC Output enabled with correspondence to air velocity There are four user selectable ranges to cho...

Страница 15: ...irmware 5 When you have completed the Factory Reset Procedure navigate to the info screen from home to validate that your new firmware version was properly installed The date of the firmware release l...

Страница 16: ...he RoosterTM Sensor100 is warranted to be free from defects in workmanship and materials for a period of one 1 year from receipt Degree Controls will at its option replace or repair defective parts wi...

Отзывы: