8       

012208 

In the majority of Ronan installations, the source is 
contained in a lead-filled source holder with an ON/OFF 
mechanism. The holder is designed so that the radiation 
field is 5 mR/hr or less at a distance of 12 inches from 
the surface of the holder when it is in the OFF position. 
When the source holder is mounted on the pipe or 
vessel and turned to the ON position, the pipe walls, 
process material and mounting bracket absorb most of 
the radiation. Again, the field intensity is about 5 Mr/hr at 
a distance of 12 inches from the surface of the gage. 
Thus, a person would have to be within 12 inches of the 
gage for 20 hours per week to receive 100 mrem. A 
person would have to be within 12 inches of the gage for 
five hours per week before he would be required to have 
a personnel monitoring device, such as a film badge or a 
dosimeter. 
 
Long experience in the nuclear gaging industry with 
hundreds of gages, where the source is contained in a 
source holder, indicates that the dose received by 
operators, maintenance personnel and supervisors 
averages less than 25 mrem per week. Thus, for gages 
where the source is contained in a source holder, it is 
usually not necessary to provide any personnel with 
monitors. 
 
Whether or not monitoring devices are needed should be 
determined at the gage site at the time of installation in 
the form of an 

occupancy evaluation.

 Only those persons 

specially licensed by the NRC or an agreement state to 
install a radioactive device are qualified to make an 
occupancy evaluation. 
 
In some installations, it is impossible to mount the source 
in a source holder. In these cases the source is usually 
mounted in a source well. Installation of the source in the 
well should be done as rapidly as possible. All necessary 
equipment should be assembled prior to opening the 
shipping box containing the source. 
 
A trial installation using a dummy source is 
recommended. A dummy source can easily be fabricated 
from steel or brass using the outline drawing of the 
source supplied by Ronan Engineering. When an 
unshielded source is installed in a vessel or when it is 
wipe tested, the radiation field is usually greater than 100 
mR/hr. Thus, personnel monitoring in the form of film 
badges or dosimeters is required. A record of the  film 
badge or dosimeter reading must be kept on Form NRC-
5. Since records must be kept, the employer must 
furnish to the employee, if requested, a record of the 
employee’s radiation exposure annually and on 
termination of employment. 
 
Additional precautions are required when a gage is used 
on a vessel large enough to permit entry of personnel. 
With the source holder in the open position or when the 
source is not removed from the source well, the radiation 
field intensity inside the vessel can be high. A procedure 

must

 be established so that personnel cannot enter the 

vessel until the source holder is in the closed position or 
the source is removed from the source well. The use of  

 

padlocks on all man-way and access port covers is 
acceptable. The key or combination for the locks should 
be kept by the person responsible for radiation safety. 
 
In some cases, when the vessel or pipe is empty, the 
radiation field intensity of the outside of the pipe or 
vessel will be such that personnel monitoring is required. 
For installations using source holders, this problem can 
easily be solved by turning the source holder to the OFF 
position. For installations using sources in source wells, 
where the radiation cannot be turned off, it may be 
desirable to remove the source temporarily and return it 
to its lead shielded shipping and storage container. 
 
 

11.1 Field Intensity Calculation

 

 

The best method for determining the radiation field 
intensity is by measurement with a survey meter. 
However, the field intensity can be calculated fairly 
accurately without a survey meter.   

 
The radiation field intensity can 
be calculated from: 
 
 

 
 
D = dose rate, mR/hr 

 

mCi = millicurie value of source 

 

d = distance to source in inches 

 

k = constant 0.023 for AM-241 
                      0.5 for CS-137 
                      2.0 for CO-60 
 
Suppose that for a certain installation the estimated 
exposure time to the unshielded source is 10 
minutes at an average body-to-source distance of 
20 inches. The source is 10 millicuries of CS-137.  
 
The dose rate would be: 

And the dosage received would be: 

 
To calculate the radiation field intensity on the 
outside of the vessel with the source installed in the 
vessel wall, a set of transmission curves is needed. 
The graph in Figure 2 shows the percentage of 
transmission of CS-137 versus material, thickness 
for lead, steel, concrete and water. 
 

 
 

1000

 

X

 

2

(d)

mCi

 

k

 

 

D

=

mr/hr

 

12.5

 

 

1000

 

X

 

400

5

 

 

1000

 

X

 

2

(20)

10

 

X

 

0.5

 

 

rate

 

Dose

=

=

=

mrem

 

2.08

 

 

mR/hr

 

12.5

 

 

min/hr

 

60

min

 

10

 

 

dose

 

Total

=

=

=

Содержание X 90 Series

Страница 1: ...ries X 90 Point Level Monitors Programmable Microprocessor POINT LEVEL MONITORS SERIES X90 Instructions and Operating Manual Customer s Name ________________________ Customer s P O No ______________________ Ronan Job No ___________________________ ...

Страница 2: ...ation NRC Regulations Periodic Leakage Test Radiation Publications Abandonment and Disposal Prohibition of Operation Mounting and Start Up Well Source Holders SA 4 SA 10 SA 15 Source retraction procedure Wiping the test points Standard Sources SA 1 SA 8 Mandatory Reporting Leak Test Procedures for Sealed Sources Ronan Leak Testing Service Using the Leak Test Kit 3 3 4 4 4 4 4 4 4 4 5 5 5 5 5 6 6 6...

Страница 3: ...tract the source into the storage source holder CAUTION Any malfunction of the source holder must be immediately reported to Ronan Engineering for repair or replacement of the source holder Under no circumstances may the source holder be taken apart Should it be necessary to ship the source holder back to Ronan contact Ronan Engineering for detailed shipping instructions A Geiger Mueller tube is u...

Страница 4: ...s as a low limit switch in which case the system will produce an alarm when the liquid level falls below this low limit The liquid itself may be under pressure at high temperature or even corrosive but its characteristics will not affect the switch since the system components are outside of the vessel A time delay based on the measurement controls the interval between radiation level change and al...

Страница 5: ...cond is the Buildup Alarm enable The jumper must be set to Enabled or Disabled depending upon the application Note that if the jumper for Alarm is set to Low Level Alarm the Buildup Alarm will be disabled regardless of the jumper setting The standard or custom parameters will already have been programmed into the unit at the factory The user must then mount and wire to the switch Once a reference ...

Страница 6: ... the switch has been calibrated using the above method the operation of the control relay NORMAL and ALARM indicators should be checked by raising and lowering the process level about the set limit If all these operate as required the switch is considered calibrated 7 4 Calibration without Process Material a Switch on the power to the X90 and turn the shutter handle on the source holder to the ON ...

Страница 7: ...erials and is able to pass through several inches of steel or other dense material The ability to penetrate dense material can be used to advantage in the measurement of process variables such as density level and thickness where a change in detected radiation indicates a change in process variable Radiated energy is harmful to the human body when absorbed at an excessive rate For example a glowin...

Страница 8: ... A record of the film badge or dosimeter reading must be kept on Form NRC 5 Since records must be kept the employer must furnish to the employee if requested a record of the employee s radiation exposure annually and on termination of employment Additional precautions are required when a gage is used on a vessel large enough to permit entry of personnel With the source holder in the open position ...

Страница 9: ...9 012208 Figure 2 Radiation Transmission CS 137 for Various Materials ...

Страница 10: ...onsible for the source the license must be amended prior to the assignment of the new user 30 32 30 33 30 34 30 38 b Use of the source is usually licensed for a particular plant site If the source is transferred to a different plant site the license must be amended prior to the transfer 30 34 and 30 38 c A record of the initial radiation survey must be kept for reference 20 40 1 b d Records of the...

Страница 11: ...nnel do not work in or frequent a restricted area Form NRC 3 need not be posted 11 3 Periodic Leakage Test Information about the periodic leakage test is covered in Section 12 11 4 Radiation Publications For those who desire further information on radiation safety and the handling of radioactive material the following publications are recommended Radiation Dosimetery Hine and Brownell Academic Pre...

Страница 12: ...CGB fitting at source rod c Remove handle and attach source extender rod e Push source rod until source bottoms out in well f Tighten CGB fitting over source extender rod 11 8 3 Source retraction procedure a Loosen CGB filling b Pull source handle back until source rod locates inside source tube Source will hit positive stop inside holder c Push shutter to OFF Lock with lock pin or padlock d Remov...

Страница 13: ...the source leaks Do not touch the cotton tipped end or allow it to touch other objects as this would spread contamination if a leak is present After making the wipe test replace the cotton swab in the vial with the cotton tipped end at the bottom Replace the cap on the vial place the vial in the shipping tube and replace the top on the shipping tube Send to Ronan Engineering Company 8050 Productio...

Страница 14: ...terials issue a report based on the results and forward a copy of the results to the customer immediately 12 2 Using the Leak Test Kit For instructions on using the Leak Test Kit see section 11 8 4 for well sources or section 11 9 2 for standard sources Return the cotton swab in the sealed plastic bag prepaid via United Parcel Services Fed Ex etc Do not send via conventional mail as postal regulat...

Отзывы: