Ronan X 90 Series Скачать руководство пользователя страница 4

4       

012208 

Failure Alarm: 

Integral: 

 Included 

 Remote:  

Included 

 
Time Delay Range:  

Integral:  

0.5 to 20 seconds (10 or 

20 seconds normal); jumper-selectable 

Remote:  

.05 to 

20 seconds (10 or 20 seconds normal); jumper-
selectable 

 

Display:  

Integral:  

LED status indicator 

Remote:  

LED 

status indicator 

 

Accuracy:  

Integral 

± 1/8” 

Remote:  

± 1/8” 

 

All equipment approved by CSA. 
 
————————————————————————— 
 

3.0 THEORY OF OPERATION 

 
The Ronan Series X90 gamma switches provide an 
alarm signal whenever the radiation field intensity at the 
detector  changes by more than its predetermined value. 
The variation in the field intensity may be due to either a 
change in the level (at the limit), or a change in the 
density of the process material in the radiation beam 
path. In a typical arrangement, the X90 detects the level 
of liquid in a vessel when it has exceeded a 
predetermined limit. The level limit is defined by an 
imaginary straight line drawn from the source to the 
detector. 

 

 
 
 
 
 
 
 
 
 

Figure 1:  

Source and Detector 

 
When the feed valve is open, liquid is fed into the vessel 
and the liquid level rises. The liquid level will rise until it 
interrupts the gamma ray beam. There is then a change 
in the field intensity at the detector due to absorption by 
the process of some of the energy present in the beam, 
and the instrument puts out an alarm signal in the form 
of a relay contact changeover. This signal may be used 
to close the feed valve and/or activate an annunciator. 
As the liquid is used from the vessel, the level will fall, 
thus again exposing the detector to the gamma radiation 
and the feed valve will be opened when the contacts 
revert to normal. 
 
This is an example of a high limit switch, i.e., the process 
level  is  not  allowed   to  rise  above  a  certain  limit.   

Conversely, the source and the detector may be lowered 
to below the liquid level and the system functions as a 
low limit switch, in which case the system will produce an 
alarm when the liquid level falls below this low limit. The 
liquid itself may be under pressure at high temperature 
or even corrosive, but its characteristics will not affect 
the switch, since the system components are outside of 
the vessel. 
 
A time delay based on the measurement controls the 
interval between radiation level change and alarm. A 
hysteresis band about the required level prevents 
fluttering and spurious alarm situations. 
 

3.1 Optional Gaging Configurations 
 

3 . 1 . 1   R e m o t e   P r o b e :

 

T h e  

R o n a n                  

Remote Probe (located on vessel) with the X90-301 
switch electronics may be located in a general 
purpose or Division II location. This is the most 
widely used configuration.  
 

3.1.2  Integral Switch:  

The Ronan Model  Integral 

Switch in an explosion-proof housing is usually 
specified when the switch electronics cannot be 
located in a general purpose or Division II area. It 
may be used when the vessel is easily accessible 
for calibration and maintenance. 
 

————————————————————————— 
 

4.0 FUNCTIONAL OPTIONS 
 

 

4.1 High-Limit Process Alarm 

 

The Process alarm relay will change over when the 
level of process material rises above the set limit. 

 

4.2 Low-Limit Process Alarm 
 

The process alarm relay will change over when the 
process level falls below the set limit. 

 

4.3 Failure Alarm 
 

In the case of the high-limit switch, under normal 
conditions the detector receives full radiation and 
produces an output. This output will fail, producing 
an alarm when either the radiation is interrupted by 
the process level rising to above the set limit or the 
detector or its power supply fails. However, if the 
detector or power supply fails, a failure alarm will 
be actuated, indicating  the  problem  is in  the unit. 
If the detector output does not indicate a minimum 
background  value (due to a malfunction in the 
detector or its power supply) the process (and 
failure, if used) relay automatically changes over, 
indicating an alarm condition. 
 
The failure alarm will occur anytime the output from 

 
 

Содержание X 90 Series

Страница 1: ...ries X 90 Point Level Monitors Programmable Microprocessor POINT LEVEL MONITORS SERIES X90 Instructions and Operating Manual Customer s Name ________________________ Customer s P O No ______________________ Ronan Job No ___________________________ ...

Страница 2: ...ation NRC Regulations Periodic Leakage Test Radiation Publications Abandonment and Disposal Prohibition of Operation Mounting and Start Up Well Source Holders SA 4 SA 10 SA 15 Source retraction procedure Wiping the test points Standard Sources SA 1 SA 8 Mandatory Reporting Leak Test Procedures for Sealed Sources Ronan Leak Testing Service Using the Leak Test Kit 3 3 4 4 4 4 4 4 4 4 5 5 5 5 5 6 6 6...

Страница 3: ...tract the source into the storage source holder CAUTION Any malfunction of the source holder must be immediately reported to Ronan Engineering for repair or replacement of the source holder Under no circumstances may the source holder be taken apart Should it be necessary to ship the source holder back to Ronan contact Ronan Engineering for detailed shipping instructions A Geiger Mueller tube is u...

Страница 4: ...s as a low limit switch in which case the system will produce an alarm when the liquid level falls below this low limit The liquid itself may be under pressure at high temperature or even corrosive but its characteristics will not affect the switch since the system components are outside of the vessel A time delay based on the measurement controls the interval between radiation level change and al...

Страница 5: ...cond is the Buildup Alarm enable The jumper must be set to Enabled or Disabled depending upon the application Note that if the jumper for Alarm is set to Low Level Alarm the Buildup Alarm will be disabled regardless of the jumper setting The standard or custom parameters will already have been programmed into the unit at the factory The user must then mount and wire to the switch Once a reference ...

Страница 6: ... the switch has been calibrated using the above method the operation of the control relay NORMAL and ALARM indicators should be checked by raising and lowering the process level about the set limit If all these operate as required the switch is considered calibrated 7 4 Calibration without Process Material a Switch on the power to the X90 and turn the shutter handle on the source holder to the ON ...

Страница 7: ...erials and is able to pass through several inches of steel or other dense material The ability to penetrate dense material can be used to advantage in the measurement of process variables such as density level and thickness where a change in detected radiation indicates a change in process variable Radiated energy is harmful to the human body when absorbed at an excessive rate For example a glowin...

Страница 8: ... A record of the film badge or dosimeter reading must be kept on Form NRC 5 Since records must be kept the employer must furnish to the employee if requested a record of the employee s radiation exposure annually and on termination of employment Additional precautions are required when a gage is used on a vessel large enough to permit entry of personnel With the source holder in the open position ...

Страница 9: ...9 012208 Figure 2 Radiation Transmission CS 137 for Various Materials ...

Страница 10: ...onsible for the source the license must be amended prior to the assignment of the new user 30 32 30 33 30 34 30 38 b Use of the source is usually licensed for a particular plant site If the source is transferred to a different plant site the license must be amended prior to the transfer 30 34 and 30 38 c A record of the initial radiation survey must be kept for reference 20 40 1 b d Records of the...

Страница 11: ...nnel do not work in or frequent a restricted area Form NRC 3 need not be posted 11 3 Periodic Leakage Test Information about the periodic leakage test is covered in Section 12 11 4 Radiation Publications For those who desire further information on radiation safety and the handling of radioactive material the following publications are recommended Radiation Dosimetery Hine and Brownell Academic Pre...

Страница 12: ...CGB fitting at source rod c Remove handle and attach source extender rod e Push source rod until source bottoms out in well f Tighten CGB fitting over source extender rod 11 8 3 Source retraction procedure a Loosen CGB filling b Pull source handle back until source rod locates inside source tube Source will hit positive stop inside holder c Push shutter to OFF Lock with lock pin or padlock d Remov...

Страница 13: ...the source leaks Do not touch the cotton tipped end or allow it to touch other objects as this would spread contamination if a leak is present After making the wipe test replace the cotton swab in the vial with the cotton tipped end at the bottom Replace the cap on the vial place the vial in the shipping tube and replace the top on the shipping tube Send to Ronan Engineering Company 8050 Productio...

Страница 14: ...terials issue a report based on the results and forward a copy of the results to the customer immediately 12 2 Using the Leak Test Kit For instructions on using the Leak Test Kit see section 11 8 4 for well sources or section 11 9 2 for standard sources Return the cotton swab in the sealed plastic bag prepaid via United Parcel Services Fed Ex etc Do not send via conventional mail as postal regulat...

Отзывы: