background image

6       

012208 

9.0 MENU VARIABLES (

accessible with hand 

held programmer) 
 

The following Menu Variables are available:  (some may be 
unavailable due to the selections of certain other options) 
 

Reference Time

: This is the time in seconds the system will 

spend averaging the incoming counts. The resultant value 
will be used to calculate all alarm trip and reset points. 

Time Constant

: Range 0 to 999 seconds. This is the count 

smoothing factor. Incoming raw counts will be filtered using 
the Time Constant value to reduce statistical noise from the 
detector. The longer the Time Constant, the more stable 
the counts, but the longer the  response time of the switch. 

Count Period: 

 .10 to 300 seconds. The period the system 

will accumulate counts before submitting the total to the 
system for filtering and alarm monitoring. 

Pre Scale Divider: 

 Range 1 to 32 in binary steps. Used in 

the event the counts exceed the system’s capabilities. 

Sigma Multiplier

: Range .1 to 9.9. This value is used to 

calculate the alarm trip point and represents a multiplier for 
one sigma of the Reference Counts. 

Hysteris

is: Range 1% to 99%. Used to calculate the reset 

or alarm point for level alarms. The Alarm/Reset Point is 
calculated based on the type of switch selected by the 
jumpers. 

Buildup Timer Start %: 

 1 to 99%. Used ONLY for the High  

Level Alarm type with Buildup Alarm enabled. It is used to 
calculate the timer start point. 

Detector High Fault  Point: 

 Range 0 to 32767 counts. The 

value here is absolute counts. From any count period 
exceed the High Fault Point, the Failure Alarm will be 
enabled. This value should be set beyond normal operating 
range. 
 
 

 

7.3  Calibration with Process Material 

 

a)  Switch on the power to the X90 and turn the 
shutter handle on the source holder to the ON 
position. Press the Start Ref button. 
 
b)  Lower the process level below the set limit 
and the system will return. 
 
After the switch has been calibrated using the 
above method, the operation of the control relay, 
NORMAL and ALARM indicators should be 
checked by raising and lowering the process level 
about the set limit. If all these operate as 
required, the switch is considered calibrated. 
 

7.4  Calibration without Process Material 

 

a) Switch on the power to the X90 and turn the 
shutter handle on the source holder to the ON 
position. Press the Start Ref button. 
b)  Turn the source OFF. 
c)  Turn the source back ON for operation. 
 

7.5  Response Time 

 

The response time of the amplifier is 0.5 to 40 
seconds, depending on the SPAN setting. 
 
 

8.0 IMPLEMENTATION 
 

Both styles of switches will accept 120-240VAC or 
24VDC power They have two (2) relays for the three 
(3) alarms: Process alarm, Buildup alarm, and Failure 
Alarm. Four (4) status LED’s will occupy the front panel 
and indicate the switch status. A set of terminals for 
the two (2) alarm relays and the remote Reference 
pushbutton will be available on each unit  also. 
 
Note that the relay outputs are wired fail-safe. The 
relays are deactivated for power-fail and alarm 
conditions. 
 
 

 
 

Status 

Fail 
LED 

Alarm 
Relay 

Rem. 
Ref 
LED 

Normal 

ON 

OFF 

OFF 

OFF 

OFF 

OFF 

OFF 

Level 
Alarm 

OFF 

OFF 

ON 

OFF 

OFF 

ON 

OFF 

Buildup 
Alarm 

OFF 

FLASH  FLASH  OFF 

ON 

ON 

OFF 

Det. 
Fail 

OFF 

ON 

OFF 

OFF 

ON 

OFF 

OFF 

Ref(after 
button 
held for 
5 sec-
onds 

OFF 

OFF 

OFF 

FLASH  OFF 

OFF 

FLASH 

Det. Fail 
during 
Refer-
ence 

OFF 

ON 

OFF 

RAPID 
FLASH 
(5 SEC) 

ON 

OFF 

RAPID 
FLASH 
(5 

Ref 
LED 

Normal 
LED 

Level 
LED 

Fail 
Relay 

out” the incoming counts to the processor module, in 
this way producing zero counts in. It is intended to be  
used only when the gage is in a  

Reference 

condi-

tion; I.e.: empty vessel with the source open. 
 
With the switch type set to Low Alarm and in the Ref-
erence condition, the switch would normally be in 
alarm. Pushing the Push-to-Test button  will clear the 
alarm. With the switch set to High Alarm and in the 
Reference condition, pressing the Push-to-Test but-
ton will cause an alarm. In each case, holding the 
Push-to-Test button down for  (2) two minutes will 
also cause a  Failure Alarm. 
 

Содержание X 90 Series

Страница 1: ...ries X 90 Point Level Monitors Programmable Microprocessor POINT LEVEL MONITORS SERIES X90 Instructions and Operating Manual Customer s Name ________________________ Customer s P O No ______________________ Ronan Job No ___________________________ ...

Страница 2: ...ation NRC Regulations Periodic Leakage Test Radiation Publications Abandonment and Disposal Prohibition of Operation Mounting and Start Up Well Source Holders SA 4 SA 10 SA 15 Source retraction procedure Wiping the test points Standard Sources SA 1 SA 8 Mandatory Reporting Leak Test Procedures for Sealed Sources Ronan Leak Testing Service Using the Leak Test Kit 3 3 4 4 4 4 4 4 4 4 5 5 5 5 5 6 6 6...

Страница 3: ...tract the source into the storage source holder CAUTION Any malfunction of the source holder must be immediately reported to Ronan Engineering for repair or replacement of the source holder Under no circumstances may the source holder be taken apart Should it be necessary to ship the source holder back to Ronan contact Ronan Engineering for detailed shipping instructions A Geiger Mueller tube is u...

Страница 4: ...s as a low limit switch in which case the system will produce an alarm when the liquid level falls below this low limit The liquid itself may be under pressure at high temperature or even corrosive but its characteristics will not affect the switch since the system components are outside of the vessel A time delay based on the measurement controls the interval between radiation level change and al...

Страница 5: ...cond is the Buildup Alarm enable The jumper must be set to Enabled or Disabled depending upon the application Note that if the jumper for Alarm is set to Low Level Alarm the Buildup Alarm will be disabled regardless of the jumper setting The standard or custom parameters will already have been programmed into the unit at the factory The user must then mount and wire to the switch Once a reference ...

Страница 6: ... the switch has been calibrated using the above method the operation of the control relay NORMAL and ALARM indicators should be checked by raising and lowering the process level about the set limit If all these operate as required the switch is considered calibrated 7 4 Calibration without Process Material a Switch on the power to the X90 and turn the shutter handle on the source holder to the ON ...

Страница 7: ...erials and is able to pass through several inches of steel or other dense material The ability to penetrate dense material can be used to advantage in the measurement of process variables such as density level and thickness where a change in detected radiation indicates a change in process variable Radiated energy is harmful to the human body when absorbed at an excessive rate For example a glowin...

Страница 8: ... A record of the film badge or dosimeter reading must be kept on Form NRC 5 Since records must be kept the employer must furnish to the employee if requested a record of the employee s radiation exposure annually and on termination of employment Additional precautions are required when a gage is used on a vessel large enough to permit entry of personnel With the source holder in the open position ...

Страница 9: ...9 012208 Figure 2 Radiation Transmission CS 137 for Various Materials ...

Страница 10: ...onsible for the source the license must be amended prior to the assignment of the new user 30 32 30 33 30 34 30 38 b Use of the source is usually licensed for a particular plant site If the source is transferred to a different plant site the license must be amended prior to the transfer 30 34 and 30 38 c A record of the initial radiation survey must be kept for reference 20 40 1 b d Records of the...

Страница 11: ...nnel do not work in or frequent a restricted area Form NRC 3 need not be posted 11 3 Periodic Leakage Test Information about the periodic leakage test is covered in Section 12 11 4 Radiation Publications For those who desire further information on radiation safety and the handling of radioactive material the following publications are recommended Radiation Dosimetery Hine and Brownell Academic Pre...

Страница 12: ...CGB fitting at source rod c Remove handle and attach source extender rod e Push source rod until source bottoms out in well f Tighten CGB fitting over source extender rod 11 8 3 Source retraction procedure a Loosen CGB filling b Pull source handle back until source rod locates inside source tube Source will hit positive stop inside holder c Push shutter to OFF Lock with lock pin or padlock d Remov...

Страница 13: ...the source leaks Do not touch the cotton tipped end or allow it to touch other objects as this would spread contamination if a leak is present After making the wipe test replace the cotton swab in the vial with the cotton tipped end at the bottom Replace the cap on the vial place the vial in the shipping tube and replace the top on the shipping tube Send to Ronan Engineering Company 8050 Productio...

Страница 14: ...terials issue a report based on the results and forward a copy of the results to the customer immediately 12 2 Using the Leak Test Kit For instructions on using the Leak Test Kit see section 11 8 4 for well sources or section 11 9 2 for standard sources Return the cotton swab in the sealed plastic bag prepaid via United Parcel Services Fed Ex etc Do not send via conventional mail as postal regulat...

Отзывы: