background image

14       

012208 

12.0  LEAK TEST PROCEDURES FOR 
          SEALED SOURCES 

 
12.1  Ronan Leak Testing Service 
 

The NRC requires that all sealed sources be tested 
for leakage at specified intervals. Only Kr-85, 
tritium and certain sources of very small activity (10 
CFR 30.18a) are exempted from leak testing. 
 
Ronan Engineering provides for users’ convenience 
and safety the “Leak Testing Service.”  The leak 
testing procedure detailed in these instructions 
applies only to the Ronan Model WK Leak Testing 
Kit supplied with the Ronan Leak Testing Service. 
The Ronan Model WK Leak Testing Kit is intended 
for use on all sealed sources that are not exempt 
from testing. 
 
This service may be purchased from Ronan on a 
contract basis for five years when wipe testing at 
six-month intervals is required or on a one-time 
basis in the case of three-year wipe test intervals. 
The service must be purchased for each individual 
source. If the service is retained, Ronan will send 
the required number of leak test kits, consisting of:   
 
a)  A cotton swab (“Q-tip”), contained in a small 
plastic sealed bag. 
 
b)  Special instructions on its use. 
 
c)  Leak test forms. 
 
After the sealed sources have been wiped by the 
customer, the kit is to be returned to Ronan for 
analysis.    Ronan    will    analyze    the   leak    
test materials, issue a report based on the results 
and forward a copy of the results to the customer 
immediately. 

 
 

12.2  Using the Leak Test Kit:

 

 

For instructions on using the Leak Test Kit see 
section 11.8.4 for well sources or section 11.9.2 for 
standard sources. 
 
Return the cotton swab in the sealed plastic bag, 
prepaid, via United Parcel Services, Fed Ex, etc. 
Do 

not

 send via conventional mail, as postal 

regulations prohibit mailing of radioactive material 
in this form. 
 
Upon receipt of the wipe, Ronan will perform a 
sensitive analysis to determine the presence of 
radioactive material. If the wipe is contamination-
free, a notice will be sent via mail that the source is 
leak-free. If radioactive material exceeding .005 
microcuries is detected on the wipe, an emergency 
notification will be sent via telegram or telephone, 
advising that the source holder must be taken out 
of service and sent back for repair. The emergency 
notification will contain detailed instructions for 
removal and shipment of source holder. 
 
Ronan will maintain records of each leak test 
analysis performed. The records will include the 
name and address of the customer; the date the 
sample was collected; the individual collecting the 
sample; the person performing the analysis; the 
date the analysis was performed; the unique 
identification of the source being tested; the 
radioactive material and mass number contained in 
the source; and the results of the test expressed in 
microcuries. 

 
 
 

 
 

Содержание X 90 Series

Страница 1: ...ries X 90 Point Level Monitors Programmable Microprocessor POINT LEVEL MONITORS SERIES X90 Instructions and Operating Manual Customer s Name ________________________ Customer s P O No ______________________ Ronan Job No ___________________________ ...

Страница 2: ...ation NRC Regulations Periodic Leakage Test Radiation Publications Abandonment and Disposal Prohibition of Operation Mounting and Start Up Well Source Holders SA 4 SA 10 SA 15 Source retraction procedure Wiping the test points Standard Sources SA 1 SA 8 Mandatory Reporting Leak Test Procedures for Sealed Sources Ronan Leak Testing Service Using the Leak Test Kit 3 3 4 4 4 4 4 4 4 4 5 5 5 5 5 6 6 6...

Страница 3: ...tract the source into the storage source holder CAUTION Any malfunction of the source holder must be immediately reported to Ronan Engineering for repair or replacement of the source holder Under no circumstances may the source holder be taken apart Should it be necessary to ship the source holder back to Ronan contact Ronan Engineering for detailed shipping instructions A Geiger Mueller tube is u...

Страница 4: ...s as a low limit switch in which case the system will produce an alarm when the liquid level falls below this low limit The liquid itself may be under pressure at high temperature or even corrosive but its characteristics will not affect the switch since the system components are outside of the vessel A time delay based on the measurement controls the interval between radiation level change and al...

Страница 5: ...cond is the Buildup Alarm enable The jumper must be set to Enabled or Disabled depending upon the application Note that if the jumper for Alarm is set to Low Level Alarm the Buildup Alarm will be disabled regardless of the jumper setting The standard or custom parameters will already have been programmed into the unit at the factory The user must then mount and wire to the switch Once a reference ...

Страница 6: ... the switch has been calibrated using the above method the operation of the control relay NORMAL and ALARM indicators should be checked by raising and lowering the process level about the set limit If all these operate as required the switch is considered calibrated 7 4 Calibration without Process Material a Switch on the power to the X90 and turn the shutter handle on the source holder to the ON ...

Страница 7: ...erials and is able to pass through several inches of steel or other dense material The ability to penetrate dense material can be used to advantage in the measurement of process variables such as density level and thickness where a change in detected radiation indicates a change in process variable Radiated energy is harmful to the human body when absorbed at an excessive rate For example a glowin...

Страница 8: ... A record of the film badge or dosimeter reading must be kept on Form NRC 5 Since records must be kept the employer must furnish to the employee if requested a record of the employee s radiation exposure annually and on termination of employment Additional precautions are required when a gage is used on a vessel large enough to permit entry of personnel With the source holder in the open position ...

Страница 9: ...9 012208 Figure 2 Radiation Transmission CS 137 for Various Materials ...

Страница 10: ...onsible for the source the license must be amended prior to the assignment of the new user 30 32 30 33 30 34 30 38 b Use of the source is usually licensed for a particular plant site If the source is transferred to a different plant site the license must be amended prior to the transfer 30 34 and 30 38 c A record of the initial radiation survey must be kept for reference 20 40 1 b d Records of the...

Страница 11: ...nnel do not work in or frequent a restricted area Form NRC 3 need not be posted 11 3 Periodic Leakage Test Information about the periodic leakage test is covered in Section 12 11 4 Radiation Publications For those who desire further information on radiation safety and the handling of radioactive material the following publications are recommended Radiation Dosimetery Hine and Brownell Academic Pre...

Страница 12: ...CGB fitting at source rod c Remove handle and attach source extender rod e Push source rod until source bottoms out in well f Tighten CGB fitting over source extender rod 11 8 3 Source retraction procedure a Loosen CGB filling b Pull source handle back until source rod locates inside source tube Source will hit positive stop inside holder c Push shutter to OFF Lock with lock pin or padlock d Remov...

Страница 13: ...the source leaks Do not touch the cotton tipped end or allow it to touch other objects as this would spread contamination if a leak is present After making the wipe test replace the cotton swab in the vial with the cotton tipped end at the bottom Replace the cap on the vial place the vial in the shipping tube and replace the top on the shipping tube Send to Ronan Engineering Company 8050 Productio...

Страница 14: ...terials issue a report based on the results and forward a copy of the results to the customer immediately 12 2 Using the Leak Test Kit For instructions on using the Leak Test Kit see section 11 8 4 for well sources or section 11 9 2 for standard sources Return the cotton swab in the sealed plastic bag prepaid via United Parcel Services Fed Ex etc Do not send via conventional mail as postal regulat...

Отзывы: