Commissioning
R&S
®
TS-PSU12
26
User Manual 1504.4524.12 ─ 05
6 Commissioning
6.1 Installation of the PSU power module
To install the R&S TS-PSU12 module, proceed as follows:
●
Run down and power off the TSVP
To facilitate optimal heat dissipation, if multiple R&S TS-PSU12 modules are used,
they should be installed distributed in the R&S CompactTSVP / R&S PowerTSVP
housing.
●
Remove the front panel from the rear side of the TSVP chassis by slackening off
the screws
Damaged backplane due to bent pins
Bent pins may result in permanent damage to the backplane.
Check the backplane connector for bent pins!
Any pins that are bent must be straightened!
When module is connected, it must be guided with both hands and carefully pressed
into the backplane connector.
●
Press in the module applying moderate pressure.
●
Tighten the upper and lower screws on the front plate of the module.
Install the PSU RIO module as described in
Chapter 6.2, "Installation of the R&S TS-
6.2 Installation of the R&S TS-PDC module
To install the plug-in module, proceed as follows:
●
The R&S TS-PSU12 module must have been installed beforehand.
●
Select the corresponding rear I/O slot for the module TS-PSU12.
●
Remove the front panel from the rear side of the TSVP chassis by slackening off
the screws
Installation of the R&S TS-PDC module