Remote Control Commands
R&S
®
SGU100A
179
User Manual 1176.7726.02 ─ 09
11.17
STATus Subsystem
This system contains the commands for the status reporting system. See also
ter A.5, "Status Reporting System"
on page 215 for detailed information.
on page 147 has no effect on the status registers.
Value ranges
●
Queries return the current value of the respective register, which permits a check of
the device status.
Return values: A decimal value in the range 0 to 32767 (=2
15
-1)
●
The configuration commands set the respective register thus determining which
status changes of the R&S
SGU cause the status registers to be changed.
Setting values: A decimal value in the range 0 to 32767 (=2
15
-1)
..................................................................................... 179
...........................................................................................179
...........................................................................................180
.................................................................................... 180
.....................................................................................180
:STATus:QUEStionable:CONDition
.................................................................................. 181
.......................................................................................181
...................................................................................... 181
:STATus:QUEStionable:NTRansition
................................................................................ 181
:STATus:QUEStionable:PTRansition
................................................................................ 182
................................................................................................182
:STATus:OPERation:CONDition?
Quieries the content of the CONDition part of the STATus:OPERation register.
This part contains information on the action currently being performed in the instru-
ment. The content is not deleted after being read out because it indicates the current
hardware status.
Return values:
<Condition>
string
Example:
:STATus:OPERation:CONDition?
Usage:
Query only
:STATus:OPERation:ENABle
<Enable>
Sets the bits of the ENABle part of the STATus:OPERation register. This setting deter-
mines which events of the Status-Event part are forwarded to the sum bit in the status
byte. These events can be used for a service request.
Parameters:
<Enable>
string
STATus Subsystem