![Rohde & Schwarz FPC Series Скачать руководство пользователя страница 336](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824336.webp)
Remote Control
R&S
®
FPC
333
User Manual 1178.4130.02 ─ 07
Parameters:
<RefPosition>
<numeric value> (integer only)
Range:
0 to 10
*RST:
8
Example:
//Define 5th grid line as the reference position for phase format
MEAS:MODE REFL
MEAS:FORM PHAS
DISP:PHAS:REF:POS 5
Manual operation:
See
DISPlay<n>:PHASe:UNWRap
<State>
This command removes the restriction limiting the value range to +/- 180°.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Unwrap phase
DISP:PHAS:UNWR ON
Manual operation:
See
DISPlay<n>:PHASe:Y:SCALe
<DisplayRange>
This command defines the range of the y-axis for the phase measurement format.
Prerequisites for this command
●
Select reflection measurement (
).
●
).
Suffix:
<n>
.
irrelevant
Parameters:
<DisplayRange>
<numeric value>
The number you enter is rounded up to the next possible display
range. For example, if you enter 300°, the R&S
FPC automati-
cally sets the display range to 360°.
Range:
90 to 100000
*RST:
360
Default unit: deg
Example:
//Define display range for phase
MEAS:MODE REFL
MEAS:FORM PHAS
DISP:PHAS:Y:SCAL 180
Remote Commands of the VNA Application