Remote Control
R&S
®
FPC
239
User Manual 1178.4130.02 ─ 07
Prerequisites for this command
●
Turn on TOI measurement (
CALCulate<n>:MARKer<n>:FUNCtion:TOI[:
Suffix:
<n>, <m>
.
irrelevant
Return values:
<TOI>
<numeric value>
Default unit: dBm
Example:
//Query TOI result
CALC:MARK:FUNC:TOI ON
CALC:MARK:FUNC:TOI:SEAR
CALC:MARK:FUNC:TOI:RES?
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:TOI:SEARchsignal
<Once>
This command searches for signals to determine the TOI.
Prerequisites for this command
●
Turn on TOI measurement (
CALCulate<n>:MARKer<n>:FUNCtion:TOI[:
Suffix:
<n>, <m>
.
irrelevant
Setting parameters:
<Once>
ONCE
Example:
//Search for signals
CALC:MARK:FUNC:TOI ON
CALC:MARK:FUNC:TOI:SEAR
Usage:
Setting only
Manual operation:
See
CALCulate<n>:MARKer<n>:FUNCtion:TOI[:STATe]
<State>
This command turns the third order intermodulation measurement on and off.
Suffix:
<n>, <m>
.
irrelevant
Remote Commands of the Spectrum Application