The Spectrum Application
R&S
®
FPC
102
User Manual 1178.4130.02 ─ 07
2. Select the "Attenuator" menu item to open the attenuator menu.
3. The R&S
FPC provides several attenuation modes.
● "Manual Att <x> dB"
Defines the attenuation manually in 5
dB steps. The range is from 0
dB to
40
dB.
Manual attenuation is indicated by a blue dot in front of the "Att:" display in the
status bar.
Figure 14-9: Indicator of manual attenuation
● "Auto Low Distortion"
Selects an attenuation and gain that results in a good signal-to-noise ratio.
Thus, it reduces the display of spurious products. It also leads to an increased
display of the inherent noise (because of a higher attenuation).
See
for the dependencies between reference level and applied
attenuation.
● "Auto Low Noise"
Selects an attenuation and gain that increases the sensitivity of the R&S
FPC.
This mode gives a better impression of spurious products and is thus useful to
analyze signals whose level is near the noise level. However, the signal-to-
noise ratio in general can deteriorate.
See
for the dependencies between reference level and applied
attenuation.
The current attenuation is indicated in the measurement settings.
Table 14-1: Dependencies of the selected reference levels and attenuation
Reference level
Preamplifier OFF
RF attenuation
Preamplifier ON
RF attenuation
Low noise
Low distortion
Low noise
Low distortion
≤-30dBm
0 dB
0 dB
0 dB
0 dB
-29 dBm to -25 dBm
0 dB
0 dB
0 dB
5 dB
-24 dBm to -20 dBm
0 dB
0 dB
0 dB
10 dB
-19 dBm to -15 dBm
0 dB
5 dB
5 dB
15 dB
-14 dBm to -10 dBm
0 dB
10 dB
10 dB
20 dB
-9 dBm to -5 dBm
5 dB
15 dB
15 dB
25 dB
-4 dBm to 0 dBm
10 dB
20 dB
20 dB
30 dB
1 dBm to 5 dBm
15 dB
25 dB
25 dB
35 dB
6 dBm to 10 dBm
20 dB
30 dB
30 dB
40 dB
11 dBm to 15 dBm
25 dB
35 dB
35 dB
40 dB
16 dBm to 20 dBm
30 dB
40 dB
40 dB
40 dB
Amplitude Configuration