Remote Control
R&S
®
FPC
236
User Manual 1178.4130.02 ─ 07
Note
:
●
To turn on the harmonic disortion measurement, use
MARKer<m>:FUNCtion:HARMonics[:STATe]
●
To turn on the AM modulation depth measurement, use
MARKer<m>:FUNCtion:MDEPth[:STATe]
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<Measurement>
CPOWer
Selects the channel power measurement.
OBANdwidth | OBWidth
Selects the occupied bandwidth measurement (OBW).
TDMA
Selects the TDMA power measurement.
Example:
//Select the channel power measurement
CALC:MARK:FUNC:POW:SEL CPOW
Manual operation:
See
See
See
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe]
<State>
This command turns a power measurement on and off.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<State>
ON | 1
Turns on the power measurement.
OFF | 0
Performs a basic spectrum measurement.
*RST:
OFF | 0
Example:
//Turn on OBW measurement
CALC:MARK:FUNC:POW:SEL OBW
CALC:MARK:FUNC:POW ON
//Return to basic spectrum measurement
CALC:MARK:FUNC:POW OFF
Manual operation:
See
See
See
See
Remote Commands of the Spectrum Application