Command Reference
R&S
®
ZNL/ZNLE
698
User Manual 1178.5966.02 ─ 07
Parameters:
<Conductance>
Conductance G<Cmp> for the specified circuit model.
Range:
-1kS to 1 kS.
Increment: 1 pS (1E-12 S)
*RST:
0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel>
STSG | SGST | GSSL | LSSG | GSSG | SLGS | SGLS | SGGS
Possible circuit models (character data); see
Manual operation:
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:L<Cmp>
<CircuitModel>, <Inductance>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:L<Cmp>?
<CircuitModel>
Specifies the inductance values L1, L2, L3 in the different circuit models for balanced
port embedding.
In the query form, the <Inductance> parameter must be omitted. The command returns
the inductance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
<Cmp>
Number of inductance in circuit model. The total number of
inductances depends on the selected circuit model.
Parameters:
<Inductance>
Inductance L<Cmp> for the specified circuit model.
Range:
-1H to 1 H.
Increment: 1 pH (1E-12 H)
*RST:
1 nH (1E-9 H)
Default unit: H
Parameters for setting and query:
<CircuitModel>
STSL | SLST | CSSL | LSSC | LSSL | SLCS | SCLS | SLLS |
GSSL | LSSG | SLGS | SGLS
Possible circuit models (character data); see
Example:
*RST; :SOUR:LPOR1 1,2; :CALC:TRAN:VNET:BAL:EMB:
PAR:L1? CSSL
Create a balanced port and query the default inductance L1 for
the Serial Cs, shunt L circuit model. The response is 1E-009 (1
nH).
CALC:TRAN:VNET:BAL:EMB:PAR:L1 CSSL, 2.2E-9
Increase the inductance to 2.2 nH.
Manual operation:
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH