VNA GUI Reference
R&S
®
ZNL/ZNLE
290
User Manual 1178.5966.02 ─ 07
only "Start Frequency" and "Stop Frequency" can be set.
Note:
For segmented frequency sweeps the start and stop frequencies and the num-
ber of sweep points are defined per segment. See
Chapter 9.5.2.2, "Define Segments
Remote command:
[SENSe<Ch>:]FREQuency:STARt
[SENSe<Ch>:]FREQuency:STOP
[SENSe<Ch>:]FREQuency:CENTer
[SENSe<Ch>:]FREQuency:SPAN
SYSTem:FREQuency?
(query frequency range of the network analyzer)
Number of Points
Sets the total number of measurement points for
This value can also be set on the "Sweep Params" tab of the "Sweep"softtool (see
Remote command:
Start Time / Stop Time
Defines the duration of a
Note:
The minimum accepted "Stop Time" may not be technically feasible. In this case,
an error message is displayed on sweep start.
Remote command:
CW Frequency
Sets the fixed frequency for
Remote command:
[SENSe<Ch>:]FREQuency:FIXed
[SENSe<Ch>:]FREQuency[:CW]
SOURce<Ch>:FREQuency<PhyPt>:FIXed
SOURce<Ch>:FREQuency<PhyPt>[:CW]
9.2.2
Power Tab
The "Power" tab provides settings related to transmit and receive power. It is also dis-
played on the "Bw Avg Power" softtool.
Stimulus Softtool