Getting Started
R&S
®
ZNL/ZNLE
67
User Manual 1178.5966.02 ─ 07
5.3.5
Activating Additional Channel Setups
The R&S
ZNL/ZNLE features multiple channel setups, i.e. you can define several mea-
surement configurations in parallel and then switch between the channel setups auto-
matically to perform the measurements sequentially. We will demonstrate this feature
by activating additional channel setups for a different frequency range, a Spectrum
measurement, and an I/Q analysis (note that the latter two measurements are only
available if the optional Spectrum mode is installed).
To activate additional channel setups
1. Press the [Mode] key on the front panel.
2. On the "New Channel Setup" tab of the "Mode" dialog box, tap the "VNA" button.
3. Change the frequency range for this measurement:
Set the
center frequency
to
500
MHz
and the
span
to
1
GHz
.
4. Press the [Mode] key on the front panel.
5. On the "New Channel Setup" tab of the "Mode" dialog box, tap the "Spectrum" but-
ton.
Figure 5-8: Adding a new channel setup
6. Create a new channel setup for I/Q analysis:
a) Press the [Mode] key.
b) Tap the "IQ Analyzer" button to activate a channel setup for the I/Q Analyzer
application.
c) Tap the "Display Config" softkey to activate the SmartGrid mode.