CALCulate<Chn>:LIMit:SEGMent<Seg>:TYPE
<LimLineType>
Selects the limit line type for a limit line segment. This can be done before or after
defining the stimulus and response values of the segment, however, a segment must
be created first to enable this command (e.g
CALC:LIM:DATA
).
Note:
The type command overwrites the
settings
and is overwritten by them. It is not affected by the other commands in the
CALCulate<Chn>:LIMit...
subsystem defining stimulus and response values of
limit lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
Parameters:
<LimLineType>
LMIN | LMAX | OFF
Limit line type
Range:
LMAX (upper limit line segment), LMIN (lower limit
line segment), OFF (limit check switched off, limit
line segment not deleted)
*RST:
LMAX
Example:
*RST; :CALC:LIM:UPP 0, 0
Define an upper limit line segment across the entire sweep
range, using a constant upper limit of 0 dBm.
CALC:LIM:SEGM:TYPE LMIN
Turn the defined limit line segment into a lower limit line seg-
ment.
Manual operation:
CALCulate<Chn>:LIMit:SOUNd[:STATe]
<Boolean>
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - Fail beep on or off.
*RST:
OFF
Example:
CALC:LIM:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
Manual operation: