Main Applications
Setting up a Run for Purification of Cellular RNA from Cultured Cells
45
B
B
Defining the Parameters in the Software
To create the order
In the
Overview
tab click
. The
Orders
sub-tab opens.
Above the
Test Parameter
area, select
Purification
from the order type list.
Select the
Cellular RNA LV
kit from the
MagNA Pure Kit Name
list. The
Purification Protocol
list is dis-
played.
Select the
Cellular RNA LV
protocol from the
Protocol
list. Additional options are displayed.
In the
Volume
group, select 200 in the
Sample
list.
In the
Volume
group, select 100 in the
Elution
list.
Select
MP96 Output Plate
from the
Target Plate
list.
Specify the sample data in the sample table with the names and positions of all samples placed in the
source plate. You can:
Enter new sample names from scratch.
Scan the barcodes of sample tubes with the handheld barcode scanner.
Load an existing sample template and modify the sample data if necessary.
In the local tool bar the
button is enabled.
After entering the parameters and the samples, the
Order
sub-tab is displayed as follows:
Click the
button. The
Save Order
dialog box opens.
Enter the file name of the order.
Click
Save
. The dialog box closes and the order file is saved in the database. The
button is en-
abled.
Содержание MagNA Pure 96
Страница 1: ...MagNA Pure 96 System Customer Training Guide Version 1 0 Software Version 2 0 December 2010...
Страница 2: ......
Страница 4: ...MagNA Pure 96 System Customer Training Guide V1 0 Table of Contents 4...
Страница 26: ...MagNA Pure 96 System Customer Training Guide V1 0 A A 26...
Страница 28: ...MagNA Pure 96 System Customer Training Guide V1 0 28 B B...
Страница 42: ...MagNA Pure 96 System Customer Training Guide V1 0 B B 42...
Страница 66: ...MagNA Pure 96 System Customer Training Guide V1 0 B B 66...
Страница 68: ...MagNA Pure 96 System Customer Training Guide V1 0 C C 68...
Страница 71: ......