cobas c
111 Installation Manual
04906900001
INSTALLATION-MANUAL
cobas c
111
Version 3.0
Page 13 of 43
4.3
Technical specifications
Software and data
handling
Operating System
LINUX
VX Works
Memory DM- Module
PXA 255
32 MB Flash ROM
32 MB DRAM
Memory RT-CPU
PXA 255
32 MB Flash ROM
32 MB DRAM
2 MB SRAM
Communication between
DM and RT
10/100 BaseT Ethernet
(for internal use only)
Communication between
RT-CPU and µC boards
TegiLink Bus (RS485)
Mass Storage
internal
Flash ROM
external (data share)
USB Memory Stick
Interfaces (external)
2 x USB 1.1/2.0
Memory stick, Modem
2 x RS-232
- Host
- Barcode Reader
Display
5.7 in. color LCD, (¼ VGA -
320 x 240 pixels); passive
matrix touch screen.
Printer
Internal thermal printer (112
mm paper width)
Samples
Sample handling
Direct tube placement
manually by user
Incubation time
Max. 12 min. / 40 cycles
Typical time to result
ABS: 4 min. to 12 min.
ISE: 90 sec. (Na, K, Cl)
Instrument cycle time
18 sec.
Reagent bottles
Content
20 ml maximum
Identification
Barcode 2-d,
Format PDF417
Number of tests
50-200 depending of test
Calibration
Use of Roche test-specific
calibrator fluids
CFAS, ISE Calibrator, etc.
Use of recommended QC
material
PreciNorm, PreciPath, etc.
Cuvette handling
6 segments of 10 cuvettes
each. Single use only
Manual handling of cuvette
segments during standby.
Incubation Temperature
37°C
±
0.2°C
Measuring Unit
Absorbance photometer
20 W halogen lamp
12 Wavelengths
340 nm ... 659 nm
Sensor
Photosensitive diode array