cobas c
111 Installation Manual
04906900001
INSTALLATION-MANUAL
cobas c
111
Version 3.0
Page 14 of 43
ISE
Ion-Selective Electrode
Indirect measurement
Sample volume
15 µl (Dilution 1:6)
OMNI
©
-Style Electrodes
Na, K, Cl, Reference
Physical
dimensions (max.)
Width /(with ISE)
590 mm / (720 mm)
Depth / (with ISE)
550 mm / (550 mm)
Height / (with ISE)
480 mm / (480 mm)
Weight / (with ISE)
32 kg (35 kg)
Packing
dimensions
Width
850 mm
Depth
700 mm
Height
850 mm
Weight
58 kg
Power require-
ments w/o ISE
Line voltage
100 – 250 V~
4 A max.
Line frequency
47 – 63 Hz
Power consumption
250 VA
This unit must be con
This unit must be con
This unit must be con
This unit must be connected to a grounded mains outlet!
nected to a grounded mains outlet!
nected to a grounded mains outlet!
nected to a grounded mains outlet!
Insulation coordination
Category II (IEC 61010-1)
Main Fuse
T6.3A H 250V
Secondary Fuses (4x)
T3.15A 250V
Battery
Replace Battery with type
Sonnenschein SL360 only
Not to be replaced by user!
ISE unit Data
(Option)
External Power Supply
UL-listed QQGQ/7 or
EPBU/7
Input Voltage:
100 – 240 V~
50 - 60 Hz
Output Voltage:
19 – 24 Vdc, Min. 4 A
Weight
3 kg
Insulation coordination
Category II (IEC 61010-1)
Power consumption
70 VA max.
Environmental
conditions
Temperature/Humidity
Transport and Storage
-25 °C to 60 °C
10-95% rH, non condensing
Temperature/Humidity
Operating
15-32°C,
30-80% rH, non condensing
Pollution
Degree 2
(IEC 61010-1)
Altitude
2000m above sea level