5
• Testing a Blood Sample
82
5.3
Performing a test
1
Check the battery level.
2
Check that the date and time are correct. Correct
any wrong entries as described in Chapter 4,
If a lockout (
Operator
or
QC Lockout
) is displayed instead
of the
Patient Test
button, you must run a liquid quality
control test before you can perform a patient test (refer to
Control Testing and Quality Control
meter is in lockout status, a test cannot be performed.
The next steps depend on whether you work
with
or
without
patient lists.
If the
Patient Test
button is available, but a lockout is
displayed, a test can only be performed as a STAT test (if
this function is enabled and if there are still STAT tests
available).
For more details on STAT test configuration see pages 71
■
If the battery icon turns red (one bar left), there may
not be enough power left for another test.
■
If there are no bars left in the battery icon, you cannot
perform any more tests. Power the meter off using the
button.
In both cases, restore power by inserting new batter-
ies, recharging the battery pack, or using the power
adapter.
Main Menu
09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Main Menu
09:15 am
04/19/2012
Control Test
Review Results
Setup
QC Lockout
Logout
Main Menu
09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Logout
QC!
Содержание CoaguChek XS Pro
Страница 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Страница 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Страница 8: ...8 This page intentionally left blank ...
Страница 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Страница 110: ...7 Review Results 110 This page intentionally left blank ...
Страница 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Страница 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Страница 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Страница 150: ...B Appendix 150 This page intentionally left blank ...
Страница 154: ...154 This page intentionally left blank ...
Страница 155: ......