9
• Maintenance and Care
124
9.3
Cleaning/disinfecting the exterior (meter housing)
When to clean and disinfect the
meter
■
The exterior of the meter should be cleaned and
disinfected before performing a test on a new
patient.
What to clean and disinfect
The following parts of the meter may be cleaned/
disinfected:
■
The area around the test strip guide
■
The meter display (touchscreen)
■
The meter housing (entire meter surface - front and
back)
■
The test strip guide and test strip guide cover
If you notice any signs of deterioration after cleaning or
disinfecting your meter, stop using it and contact Roche
Care Center at 1-877-273-3433 for assistance.
Ensure that the blue test strip guide cover remains tightly
closed while cleaning or disinfecting the exterior housing.
NOTICE
Do not allow liquid to enter the test strip guide or pooling
of liquid on the touchscreen. If liquid does get into the
test strip guide, immediately dry the components with a
dry cloth.
If solution is allowed to collect in any
meter opening, severe damage to the system can
occur.
Содержание CoaguChek XS Pro
Страница 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Страница 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Страница 8: ...8 This page intentionally left blank ...
Страница 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Страница 110: ...7 Review Results 110 This page intentionally left blank ...
Страница 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Страница 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Страница 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Страница 150: ...B Appendix 150 This page intentionally left blank ...
Страница 154: ...154 This page intentionally left blank ...
Страница 155: ......