Control Testing and Quality Control •
6
97
7
Apply gentle pressure to the reservoir to transfer
the entire contents of the dropper to the bottle.
Make sure that the dropper does not come into
contact with the dried control plasma.
8
Close the bottle again.
9
Make sure the dropper is at hand for the next steps
in the liquid quality control test.
10
Swirl the bottle using a circular motion to com-
pletely dissolve all of the control plasma inside.
Do
not shake the bottle or turn it on its side.
Doing
so can cause components in the control plasma to
stick to the sides of the bottle. Please refer to the
control solution package insert.
The control solution is now ready to be applied to the test
strip.
The control solutions may be reconstituted (mixed) after
removal from the refrigerator. The resulting solution may
be used
up to 30 minutes
after reconstitution.
Содержание CoaguChek XS Pro
Страница 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Страница 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Страница 8: ...8 This page intentionally left blank ...
Страница 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Страница 110: ...7 Review Results 110 This page intentionally left blank ...
Страница 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Страница 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Страница 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Страница 150: ...B Appendix 150 This page intentionally left blank ...
Страница 154: ...154 This page intentionally left blank ...
Страница 155: ......