4 Operation
Manual TME400-VMF · EN07 · December, 7
th
2020
55
Caution
It is necessary to remove the seals, particularly the seal over the calibration
button in order to press the calibration button (see
Figure 8: Position of the
calibration button
).
The TME400 must only be used for custody transfer with unbroken seal. Re-
moval or damage to seals normally entails considerable expenses!
Re-application of seals must only be carried out by an officially recognized
inspection authority or calibration officials!
The principle of the programming is shown based on the example of changing the
output pulse factor:
I.
Move with the arrows (
) to the position: A11
II.
Activate the calibration button (see
Figure 8: Position of the calibration button
)
III.
The blinking "INPUT" text appears above the displayed value in the display
view.
IV.
Press
for more than 2 seconds
V.
The value begins to blink at a position
VI.
With the
and
arrows, you can now increase or decrease the value at
this position. For the values, after the "0", you also have "-1" available in order
to enter negative values, if necessary.
VII.
With the
and
arrows, you can move to a different position of the
value and change it – as described in the point above.
VIII.
An additional position is added when you move with the
and
before
the displayed number.
For example, only the units digit is displayed. If you move in front of it, you will
also have the tens position available as an entry.
IX.
By pressing and holding the "right" button
, the position of the decimal
point is changed. After pressing and holding, the decimal point is inserted after
the blinking digit.
X.
By pressing and holding the "left" button
, the entry can be canceled. If a
change and/or entry is necessary, the entry must be restarted.
XI.
When you have finished making an entry, you confirm it by briefly pressing
.
XII.
A plausibility check takes place and the result is displayed immediately.
XIII.
If this check shows an implausible entry, "rAnGE" will be shown briefly in the
display and the display jumps back to the original value.
XIV.
If this check shows a plausible entry, "Good" will be shown briefly in the dis-
play and the value is adopted as a new value.
XV.
Now you can – if necessary – change other parameters.
Содержание TME400-VM
Страница 1: ...Operating Manual Turbine Meter TME400 VM VMF Stand 07 12 2020 Version 07 Firmware 1 06...
Страница 47: ...2 Installation Manual TME400 VMF EN07 December 7th 2020 41 Ex version...
Страница 48: ...2 Installation Manual TME400 VMF EN07 December 7th 2020 42 Ex version with current module...
Страница 49: ...2 Installation Manual TME400 VMF EN07 December 7th 2020 43 NON Ex version...
Страница 112: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 106...
Страница 113: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 107...
Страница 114: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 108...
Страница 115: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 109...
Страница 116: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 110...
Страница 117: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 111...
Страница 118: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 112...
Страница 119: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 113...
Страница 120: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 114...
Страница 121: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 115...
Страница 122: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 116...
Страница 123: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 117...
Страница 124: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 118...
Страница 125: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 119...
Страница 126: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 120...
Страница 127: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 121...
Страница 128: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 122...
Страница 129: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 123...
Страница 130: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 124...
Страница 131: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 125...
Страница 132: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 126...
Страница 133: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 127...
Страница 134: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 128...
Страница 135: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 129...
Страница 136: ...APPENDIX Manual TME400 VMF EN07 December 7th 2020 130...