background image

 

18 • 65-2443 PPM Hydrogen Transmitter

Returning to Normal Operation

1. Remove the voltmeter leads from the amplifier test points.

2. Unscrew the calibration cup from the detector

NOTE:

For convenience, leave the regulator and calibration cup connected by the humidifer tube.

3. Secure the junction box cover to the junction box.
4. When the controller display reading falls below the alarm setpoints, return the controller to normal 

operation.

5. Verify that the controller display reading decreases and stabilizes at 0 ppm.
6. Store the components of the calibration kit in a safe and convenient place.

Parts List

Table 6 lists replacement parts and accessories for the ppm hydrogen transmitter.

Table 6:Parts List

Part Number

Description

18-0400RK

Junction box with cover

33-2001RK-01

Humidifier tube w/3/16” tubing on ends, for calibration kit

57-1070-01

Amplifier (specify target gas and detection range)

61-0160RK

PPM hydrogen detector

65-2443

PPM hydrogen transmitter (includes detector and amplifier)

71-0542

65-2443 PPM Hydrogen Transmitter Operator’s Manual 

(this document)

81-0000RK-01

Calibration cylinder, 1,000 ppm hydrogen in air, 34 liter steel

81-0000RK-03

Calibration cylinder, 1,000 ppm hydrogen in air, 103 liter

81-0000RK-21

Calibration cylinder, 2,000 ppm hydrogen in air, 34 liter steel

81-0000RK-23

Calibration cylinder, 2,000 ppm hydrogen in air, 103 liter

81-1050RK

Regulator, 0.5 LPM, with pressure gauge and flow control knob, for 17 liter and 34 liter steel 

calibration cylinders (cylinders with external threads)

81-1051RK

Regulator, 0.5 LPM, with pressure gauge and flow control knob, for 34 liter aluminum, 58 

liter, and 103 liter steel calibration cylinders (cylinders with internal threads)

81-1117RK

Calibration cup

81-F072RK

Calibration kit, includes regulator, humidifier tube, calibration cup, 103 liter 1,000 ppm 

hydrogen calibration cylinder, and 103 liter 2,000 ppm hydrogen calibration cylinder

81-F072RK-LV

Calibration kit, includes regulator, humidifier tube, calibration cup, 34 liter 1,000 ppm hydro

-

gen steel calibration cylinder, and 34 liter 2,000 ppm hydrogen steel calibration cylinder

Содержание 65-2443

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2443 PPM Hydrogen Transmitter Operator s Manual Part Number 71 0542 Revision P1 Released 12 29 20...

Страница 2: ...c calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the typ...

Страница 3: ...air costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRESSED OR IMPLIED AN...

Страница 4: ...up 12 Introducing Incoming Power 12 Transmitter Burn In Period 12 Setting the Zero Signal 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 13 Replacing Components of the PPM Hydrogen Transm...

Страница 5: ...en transmitter WARNING When using the 65 2443 you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65 2443 and to minimize the risk of personal injur...

Страница 6: ...hydrogen transmitter s components It is a 4 to 20 mA type detector head It consists of the ppm hydrogen detector amplifier and junction box Figure 1 Component Location Amplifier J Box Hydrogen Detecto...

Страница 7: ...lifier see below Amplifier The amplifier converts the non linear electrical output from the detector to a linear 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a...

Страница 8: ...side of the amplifier see Figure 2 The test points produce a 100 to 500 mV output that corresponds to the transmitter s 4 to 20 mA output Use the test points and a voltmeter to measure the amplifier...

Страница 9: ...the mounting site Select a site where the transmitter is not likely to be bumped or disturbed Make sure there is sufficient room to perform start up maintenance and calibration procedures Select a sit...

Страница 10: ...rough the top conduit hub of the junction box 7 Pull out the controller terminal strip and connect the three wires to the terminal strip as follows see Figure 4 Connect the positive wire to the PWR SI...

Страница 11: ...rogen Transmitter to a Controller 12 If shielded cable is used connect the cable s drain wire to an available chassis earth ground at the controller RKI controllers typically have a ground stud that c...

Страница 12: ...on the controller 5 Verify that the controller is on and operating properly Refer to the controller operator s manual Transmitter Burn In Period Once power has been applied to the transmitter it will...

Страница 13: ...obable Causes Recommended Action Fail Condition Controller indicates a fail condition The transmitter wiring is disconnected or misconnected The transmitter is malfunctioning 1 Verify that the transmi...

Страница 14: ...Figure 4 on page 11 of this manual 9 Re install the detector terminal strip into its socket NOTE When a transmitter is first powered up with a new detector its output may be high enough to cause an al...

Страница 15: ...Figure 4 on page 11 of this manual If controller leads or detector leads were removed during this procedure refer to Tables 4 and 5 below 8 Install the amplifier into the junction box with the screw l...

Страница 16: ...or the environment is not well controlled then more frequent calibration than every 6 months may be necessary Calibration This section describes how to calibrate the ppm hydrogen transmitter It includ...

Страница 17: ...rogram or disable external alarms before continuing Setting the Zero Reading The detector s zero setting is controlled by electronics and is not user adjustable Setting the Span Reading 1 Screw the re...

Страница 18: ...transmitter includes detector and amplifier 71 0542 65 2443 PPM Hydrogen Transmitter Operator s Manual this document 81 0000RK 01 Calibration cylinder 1 000 ppm hydrogen in air 34 liter steel 81 0000R...

Отзывы: