7.10 Error messages and troubleshooting
This chapter includes measures in event of malfunctions. If an error occurs that is not listed in this
table, please contact your supplier.
The following faults can only occur if a rechargeable battery is used.
Problem
Possible cause
Fault solution
Drive / controller wit-
hout function
No power supply
Establish mains connection.
Drive / controller defective
Contact the supplier / dealer.
The drives cannot be tra-
versed any more.
Temperature monitoring or system
protection has tripped.
Eliminate overload (change / remove
load).
Leave the system in idle position for
approx. 20 - 30 minutes with the mains
plug removed.
Device fuse has tripped.
Contact the supplier / dealer.
Supply line (mains and / or lifting
column / operating button) inter-
rupted.
Check the supply line and, where
applicable, reestablish the connection.
Display
Fault
Fault solution
Beeping sound
The acoustic alarm warns the user
against the complete discharging
of the rechargeable battery when
in rechargeable battery mode.
The rechargeable battery must be char-
ged by plugging in the mains plug.
Display
Meaning
LED
The rechargeable battery is being
charged.
LED
The charging procedure of the
rechargeable battery is completed.