39
Specifi cations
Applicable standards
CE marking
Input
Number of channels 2
Connector type
Microdot connectors
Supported accelerometers
Piezoelectric accelerometer
Accelerometer with integrated preamplifi er, 24 V 4 mA
Measurement modes and units
Acceleration/velocity/displacement, selectable
ACC (acceleration):
m/s
2
VEL (velocity):
mm/s
DISP (displacement): mm
Sensitivity setting
Setting range
0.100 to 0.999 in 0.001 steps
1.00 to 9.99 in 0.01 steps
10.0 to 99.9 in 0.1 steps
Units pC/(m/s
2
) (using piezoelectric accelerometer)
mV/(m/s
2
) (using accelerometer with integrated pre-
amplifi er)
Range
Seven stages
Sensitivity setting determines range.
Sensitivity setting 0.100 to 0.999
ACC (acceleration):
10, 30, 100, 300, 1000, 3000,
10000
VEL (velocity):
10, 30, 100, 300, 1000, 3000,
10000
DISP (displacement): 1, 3, 10, 30, 100, 300, 1000
Sensitivity setting 1.00 to 9.99
ACC (acceleration):
1, 3, 10, 30, 100, 300, 1000
VEL (velocity):
1, 3, 10, 30, 100, 300, 1000
DISP (displacement): 0.1, 0.3, 1, 3, 10, 30, 100