Measurement
Accelerometer setup mode
This mode serves for making accelerometer related settings for the UV-16.
Before starting a measurement, be sure to make or check the accelerometer
setting.
Setting item
Setting content
Remarks
Piezoelectric accelerom-
eter
CHARGE
Setting is made separately
for channel A and B
Accelerometer with inte-
grated preamplifi er
CCLD
Sensitivity
0.100 to 0.999 in 0.001 steps
1.00 to 9.99 in 0.01 steps
10.0 to 99.9 in 0.1 steps
Setting is made separately
for channel A and B
Setting procedure
1. Pressing the PICKUP key in measurement mode activates the ac-
celerometer setup mode.
Make the setting as required by the connected accelerometer.
•
Piezoelectric accelerometer connected: Select CHARGE
•
Accelerometer with integrated preamplifi er connected: Select
CCLD
Use the RANGE
/
keys for the respective channel to make the
setting. The indicator for the selected channel fl ashes, and the other
indicator is off.
When the setting has been made with the RANGE
/
keys for
channel A and channel B, press the PICKUP key.
CHARGE
CCLD
SENS
m/s
2
mm/s mm
Hz
kHz
CHARGE
CCLD
SENS
m/s
2
mm/s mm
HPF
LPF
Hz
kHz
HPF
LPF
A
B
Same indication as for channel A
Channel A
CHARGE/CCLD indicator
SENS indicator
Range/sensitivity readout
Channel B