19
Power Supply Connection
Using the Battery Unit BP-17
Battery life
Figures for approximate battery life when using the Battery Unit BP-17 are
given below.
These fi gures are for using three UV-16 units.
Manganese batteries
Alkaline batteries
Approx. 5 hours
Approx. 17 hours
* Continuous operation at ambient temperature of 25°C, with CHARGE setting,
normal operation mode
For instructions on how to insert batteries and other information, please refer
to the documentation of the BP-17.
Important
When the upper LED is off and the lower LED
lights up in red, you should replace the batteries
with a fresh set.
When the lower LED is fl ashing in red, correct
measurement is no longer possible.
Before replacing the batteries, be sure to set
the power switch to OFF (press
side).
When inserting batteries, take care not to mix
up "+" and "−" polarity.
Always replace all eight batteries together. Using a
mixture of old and new batteries or batteries of a dif-
ferent type can lead to malfunction and damage.
When not using the unit, remove the batteries to guard
against the risk of damage by leaking battery fl uid.
When powering a system from batteries using
the BP-17, the maximum number of UV-16 units
that can be linked is three.
Battery indicator
Power switch
Battery Unit BP-17
I
O
BP-17
BATTERY
FULL
RED:CHANGE
POWER
FLASH:EMPTY
Upper LED
Lower LED
Upper LED lit in green, lower LED lit in green
Batteries have a full voltage.
Upper LED off, lower LED lit in green
Batteries no longer have a full voltage but are still operational.
Upper LED off, lower LED lit in red
Battery voltage is low.
Replace all batteries.
Upper LED off, lower LED flashing in red
Correct measurement is not possible.
Содержание UV-16
Страница 2: ......
Страница 10: ...viii...
Страница 56: ...This product is environment friendly It does not include toxic chemicals on our policy No 52099 18 07...