33
Measurement
10% of range full-scale value
(Example: 1 m/s
2
when range setting is 10 m/s
2
)
100% of range full-scale value
(Example: 10 m/s
2
when range setting is 10 m/s
2
)
50% of range full-scale value
(Example: 5 m/s
2
when range setting is 10 m/s
2
)
Bar graph indication
The bar graph indicator has 10 segments.
The vibration waveform is averaged and the resulting value is displayed on
the bar graph, using a refresh rate of 100 milliseconds.
Output signal
The AC OUT connectors on the rear panel provide a signal according to the
acceleration/velocity/displacement setting.
The amplitude of the measured vibration waveform can be determined from
the selected range and the voltage value.
A
B
Vibration waveform amplitude at point A = Range × B (unit: V)
Example
Range setting is 10 m/s
2
Range
Voltage value B
Amplitude
10 m/s
2
5 V
50 m/s
2
(Peak)
10 m/s
2
1 V
10 m/s
2
(Peak)
10 m/s
2
0.5 V
5 m/s
2
(Peak)
10 m/s
2
0.2 V
2 m/s
2
(Peak)