26
Measurement
Operation mode
The operation modes of the UV-16 are listed below.
Operation mode
Outline
Accelerometer setup mode
Make accelerometer sensitivity settings
Setup mode
Make acceleration/velocity/displacement settings,
HPF setting, LPF settings
Measurement mode
Make measurements
Calibration (CAL) mode
Use a calibration output signal to calibrate external
equipment
Default settings
If power is supplied while holding down the SETUP key, the resume informa-
tion* is cleared, and the unit starts up with the default settings listed below.
When starting up with the default settings, the range/sensitivity readout shows
the indication
for about 10 seconds.
Default settings (common for both channels)
Input setting:
CHARGE
Sensitivity setting:
5.00
Measurement mode: Acceleration (ACC) m/s
2
Range: 100
HPF:
OFF (indication --)
LPF:
OFF (indication --)
* Resume information: All settings are automatically memorized when
power is turned off.
When power is turned on, the settings are
established again.
When power comes on, the sensitivity value
at the last power off point fl ashes for about
10 seconds on the range/sensitivity readout.
CHARGE
SENS
m/s
2
Hz
kHz
CHARGE
SENS
m/s
2
HPF
LPF
Hz
kHz
HPF
LPF
A
B
Previously selected sensitivity setting
flashes for about 10 seconds.