7 Maintenance
7-2. Regular maintenance mode
- 35 -
SD-2600
<Gas Test "2-0.0">
<Alarm Test "2-0.1">
2-0.0
Press the SET key.
Introduce the test gas and
perform the gas test.
Stop introducing the test gas. When
the reading drops, press the
MENU/ESC key to cancel the test and
to go back to the original state.
To 2-0.0
WARNING
Because the contact and gas concentration output can be activated by a gas test even in the
maintenance mode depending on setting, be careful to perform the test.
WARNING
Because the contact and gas concentration output can be activated by an alarm test even in the
maintenance mode depending on setting, be careful to perform the test.
2-0.1
Press the SET key.
Alarm Test ON/OFF
(The used contact setting is "Alarm",
"Alarm or Fault")
Select either ON/OFF, and then press
the SET key.
Perform the gas alarm test by
increasing or decreasing the reading
with the ▲ or ▼ key.
Press the MENU/ESC key to cancel the
test and to go back to the original state.
To 2-0.1
▼
LED
LED