10 Product Specifications
10-1. List of specifications
- 42 -
10
Product Specifications
10-1. List of specifications
<ATEX/IECEx Specifications>
Model SD-1
Type TYPE
GP
S
Detection principle
Catalytic combustion type
Gas to be detected
Combustible gas
Concentration display
Seven-segment LED (four-digit)
Detection range
0 - 100% LEL
Display resolution
0.5% LEL
Detection method
Diffusion type
Alarm setpoint
25% LEL [standard setting value]
Power display
POWER lamp on (green)
External output
Gas concentration signal/alarm contact (gas alarm or fault alarm, gas/fault common alarm)
Alarm accuracy
(under the same conditions)
Within ±25% of the alarm setpoint value
Response time
(under the same conditions)
Within 30 seconds after providing the gas 1.6 times the alarm setpoint
Gas alarm type
One-step alarm (H)
Gas alarm display
ALM lamp on (red)
Gas alarm pattern
Auto-reset
Fault alarm/self diagnosis
System abnormalities/sensor abnormalities
Fault alarm display
FAULT lamp on (yellow)/detail display
Fault alarm pattern
Auto-reset
Alarm contact
No-voltage contact 1a, de-energized (energized in response to an alarm) or energized
(de-energized in response to an alarm)
Contact capacity
250 VAC - 0.5 A/30 VDC - 0.5 A (resistant load)
Transmission system
3-wire type analog transmission (common cable for power and signal <power, signal, common>)
Transmission specifications
4 - 20 mA DC (linear/load resistance under 300
Ω
)
Transmission cable
CVVS1.25 mm
2
or 2.0mm
2
- 3-core(when the contact is not used)
CVVS1.25 mm
2
or 2.0mm
2
- 5-core(when the contact is used)
Transmission distance
Less than 1.25km in case of CVVS 1.25mm
2
Less than 2.0km in case of CVVS 2.0mm
2
Functions
Alarm delay/suppression/zero follower
Power supply
24 VDC
±
10%
Power consumption
Maximum 3 W
Cable connecting port
Adapter A<NPT1/2> or Adapter B<NPT3/4> or
Pressure proof packing gland<G3/4> (Compatibles
φ
9.6 – 13.0 mm in outer diameter)
Initial clear
Approx. 25 seconds
Operating temperatures
-20 - +60ºC (at a constant condition)
Operating humidities
Below 95% RH (Non-condensing)
Structure
Wall mounting type
Explosion-proof structure
Flame-proof enclosures
Explosion-proof class
Ⅱ
2G Ex db
Ⅱ
C T5 Gb (ATEX) / Ex db
Ⅱ
C T5 Gb (IECEx)
Safety Integrity level
IEC 61508:2010 Part2 SIL2 capable
External dimensions
Approx. 148 (W) x 149 (H) x 88 (D) mm (projection portions excluded)