![Riken Keiki GP-1DGH Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/riken-keiki/gp-1dgh/gp-1dgh_operating-manual_1461474022.webp)
4 How to Use
4-6. How to wire
- 20 -
CAUTION
The specified bare wire length must be observed when the wire insulation is peeled off.
Improper clamping of the wire due to a shorter bare wire length may cause defective electrification or
heating.
Damaging the wire insulation due to a shorter bare wire length may cause defective electrification or
heating.
Exposing the wire due to a longer bare wire length may cause defective insulation or a short circuit.
Be careful not to break up the wire. If the wire is broken up when inserted to the terminal, this may cause
defective insulation or heating.
Compatible bar terminal
For a bar terminal, the following items are available.
Bar terminal (ferrule): Model 216 Series (manufactured by WAGO)
Crimping tool: Model VarioCrimp 4 (206-204) (manufactured by WAGO)
<How to Connect to Terminal Plate>
When cables are connected to the connectors, use the dedicated lever or a flathead screwdriver to do it as
shown below.
NOTE
<How to Use the Dedicated Handling Lever>
To check whether the wire is connected securely, pull the wire gently. (Do not pull the wire strongly.)
CAUTION
A bar terminal of the specified model must be used. Using other bar terminals invalidates the warranty of
the performance.
CAUTION
The right tools must be used.
In principal, one wire can be connected to one wiring hole.
When the wire is inserted into the driver slot by mistake, it does not contact the conductive part. This may
cause defective electrification or heating.
When the wire is inserted under the spring by mistake, it does not contact the conductive part. This may
cause defective electrification or heating.
Push the lever with your finger
to lower the spring in its inside.
While holding down the lever, insert the wire into the
(round) wiring hole until it reaches the deepest point.
Once the lever is released, the wire is secured.