5 How to Operate
5-2. Basic operating procedures
- 26 -
5-2. Basic operating procedures
Normally, the detection mode is activated after the power is turned on.
The "maintenance mode selector switch" is hereinafter called "MAINT switch".
Power-on
Preparation
for start-up
Initial clear (approx. 25 seconds)
Detection mode
Alarm test mode
(1) Slide the MAINT switch to CH1 or CH2
or
(2) Press the MODE key for over 3 seconds with the MAINT
switch at the center
Gas test mode
Simplified calibration
(zero/span adjustment)
Alarm setpoint check
Press the TEST/SET switch
Press the MODE key for
over 3 seconds
Maintenance mode
(user mode)
Maintenance mode
(regular maintenance
mode)
Press the TEST/SET key for over
3 seconds
(1) Return the MAINT switch to the
center
or
(2) Press the MODE key for over 3
seconds
WARNING
When the alarm system enters each mode from the detection mode while an alarm is activated, the
alarm contact is released.