![Rigol DSG5000 Series Скачать руководство пользователя страница 76](http://html.mh-extra.com/html/rigol/dsg5000-series/dsg5000-series_programming-manual_1460929076.webp)
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<mode>
Discrete
AUTO|EXTernal|EGATe|KEY|BUS
AUTO
Remarks
•
AUTO: selects Auto trigger. At this point, the microwave signal generator meets
the trigger condition at any time and can perform pulse modulation
continuously.
•
EXTernal: selects Ext trigger. At this point, the microwave signal generator
receives the external trigger signal input from the front-panel [TRIGGER]
connector. The instrument starts a pulse modulation each time a TTL pulse with
the specified polarity is received. To specify the polarity of the TTL pulse, use
[:SOURce][:RF]:PULM:TRIGger:EXTernal:SLOPe
command to select "Pos" or
"Neg".
•
EGATe: selects Gate trigger. At this point, the microwave signal generator
receives the external gated signal input from the front-panel [TRIGGER]
connector. The instrument starts a pulse modulation within its valid level each
time a gated signal with the specified polarity is received. To specify the polarity
of the external gated signal, use the
[:RF]:PULM:TRIGger:EXTernal:GATE:POLarity
command to select "Normal" or
"Inverse".
•
KEY: selects Key trigger. At this point, the instrument starts a pulse modulation
each time you click or tap Key.
•
BUS: selects Bus trigger. At this point, the instrument starts a pulse modulation
each time you send the
This command is invalid when "Ext" modulation source is selected.
Command System
68
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Содержание DSG5000 Series
Страница 1: ......
Страница 8: ...VI Copyright RIGOL TECHNOLOGIES CO LTD All rights reserved ...
Страница 122: ...Example N A Command System 114 Copyright RIGOL TECHNOLOGIES CO LTD All rights reserved ...
Страница 144: ......