BUS: selects Bus trigger.
If the sweep mode is "Continuous", every time you send the
command,
the instrument starts to sweep one point. If the sweep mode is "Single", you
need to send the
command first to meet the
single sweep condition. At this point, each time you send the
command,
the instrument sweeps one point and then stops after completing one sweep
period.
•
EXT: selects External trigger.
The instrument receives the trigger signal input from the front-panel [TRIGGER]
connector. If the sweep mode is "Continuous", every time the instrument
receives a TTL pulse signal with the specified polarity, the instrument starts to
sweep one point. If the sweep mode is "Single", you need to send the
command first to meet the single sweep condition. At this
point, each time the instrument receives a TTL pulse signal with the specified
polarity, it sweeps one point and then stops after completing one sweep period.
TIP
The above remarks are only valid when the trigger mode in the sweep period is met.
When executing the sweep operation, the priority of the required conditions is:
single sweep → trigger mode → point trigger mode.
Return Format
The query returns the point trigger type.
Example
:SWE:POIN:TRIG:TYPE AUTO /*Sets the point trigger type to AUTO (by
default, CH1 is selected).*/
:SWE:POIN:TRIG:TYPE? /*The query returns AUTO.*/
3.1.5.16
[:SOURce][:RF]:SWEep:SWEep:TRIGger:TYPE
Syntax
[:SOURce][:RF
<
channel>
]:SWEep:SWEep:TRIGger:TYPE
<
type>
[:SOURce][:RF
<
channel>
]:SWEep:SWEep:TRIGger:TYPE?
Description
Sets the trigger type of the sweep period for the specified channel.
Queries the trigger type of the sweep period for the specified channel.