![Rigol DSG5000 Series Скачать руководство пользователя страница 74](http://html.mh-extra.com/html/rigol/dsg5000-series/dsg5000-series_programming-manual_1460929074.webp)
Description
Sets the polarity of the external gated signal for the specified channel.
Queries the polarity of the external gated signal for the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<type>
Discrete
NORMal|INVerse
NORMal
Remarks
•
NORMal: sets the polarity of the current pulse modulation signal to "Normal".
•
INVerse: sets the polarity of the current pulse modulation signal to "Inverse".
When the trigger mode of the pulse modulation is set to "Gated", the microwave
signal generator receives the external gated signal input from the front-panel
[TRIGGER] connector. At this point, you can use this command to set the polarity of
the external gated signal. You can run the
[:SOURce][:RF]:PULM:TRIGger:MODE
command to set or query the trigger mode of the pulse modulation.
This command is invalid when "Ext" modulation source is selected. You can run the
command to set or query the pulse modulation
source.
Return Format
The query returns NORM or INV.
Example
:PULM:TRIG:EXT:GATE:POL INV /*Sets the polarity of the pulse
modulation to INV (by default, CH1 is selected).*/
:PULM:TRIG:EXT:GATE:POL? /*The query returns INV.*/
3.1.9.16
[:SOURce][:RF]:PULM:TRIGger:EXTernal:SLOPe
Syntax
[:SOURce][:RF
<
channel>
]:PULM:TRIGger:EXTernal:SLOPe
<
slope>
[:SOURce][:RF
<
channel>
]:PULM:TRIGger:EXTernal:SLOPe?
Description
Sets the slope of the valid edge of the external trigger pulse for the specified
channel.
Command System
66
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.