2-30
Marks Indicator
ķ
The horizontal time base (main time base).
ĸ
The position of the current waveform window in the memory.
Ĺ
The trigger position in the memory.
ĺ
The trigger position in the current waveform windows.
Ļ
The trigger’s horizontal offset with respect to the center of the window.
Key Points
Y-T:
The conventional oscilloscope display format. It shows that the voltage of a
waveform record (on the vertical axis) varies along with time (on the
horizontal axis).
X-Y:
XY format displays channel 1 in the horizontal axis and channel 2 in the
vertical axis.
Roll Mode:
In this mode, the waveform display rolls from right to left. No trigger
or horizontal offset control of waveforms is available during Roll Mode, and
it’s only available when set to 500ms/div or slower.
Slow Scan Mode:
This mode is available when the horizontal time base is set to
50ms or slower. In this mode, the oscilloscope acquires sufficient data for
the left part to the trigger point, then wait for trigger, when trigger occurs, it
continue to draw the rest part from the trigger point to the end of the right
side. When choosing this mode to view low frequency signals, it is
recommended that the channel coupling be set as
DC
.
Time/Div:
Horizontal scale. If the waveform acquisition is stopped (using the
Run/Stop button), the Time/Div control expands or compresses the
waveform.