Time Measurements
The DS1000B series oscilloscopes provide 12 kinds timing parameters
auto-measurements; Frequency, Period, Rise Time, Fall Time, +Width, -Width, Delay
A
ė
B , Delay A
ė
B , +Duty and –Duty, Phase A
ė
B ,Phase A
ė
B .
Figure 2-124 shows a pulse with some of the time measurement points.
Figure 2-124 The time parameters
Rise Time:
Time that the rising edge of the first pulse in the waveform takes to rise
from 10% to 90% of its amplitude.
Fall Time:
Time that the falling edge of the first pulse in the waveform takes to fall
from 90% to 10% of its amplitude.
+Width:
The width of the first positive pulse in 50% amplitude points.
-Width:
The width of the first negative pulse in the 50% amplitude points.
Delay A
ė
B :
The time delays between the channel A and channel B at the rising
edge.
Delay A
ė
B :
The time delays between the channel A and channel B at the falling
edge.
Phase A
ė
B :
The phase between the channel A and channel B at the rising edge.
Phase=(delay/source A period)
h
360
e
Phase A
ė
B :
The phase between the channel A and channel B at the falling edge.
+Duty:
+Duty Cycle, defined as +Width/Period
-Duty:
-Duty Cycle, defined as -Width/Period.
10%
50%
90%
-Width
+Width
Fall Time
Rise Time