Example 7: Triggering on a Video Signal
Test a video circuit in the DVD set. Use video trigger to obtain a stable display.
Triggering on Video Fields
To trigger on the video fields, please do as the following steps:
1
Press the MENU key in the Trigger control area to see the Trigger menu.
2
Press
Mode
to select
Video
.
3
Press
Source
to select
CH1
as trigger source.
4
Press
Polarity
to select
.
5
Press
Sync
as
Odd Field
or
Even Field
.
6
Adjust the
knob to set trigger level at the video sync pulse to get
stable trigger.
7
Turn the horizontal
knob to see a complete waveform on the screen.
Figure 3-7
Waveform display
DS1000B series triggers on the Odd field or Even field. To avoid confusion when Odd
field and Even field trigger simultaneously, choose Odd field or Even field as step 5
above.