To Set Up the Horizontal System
Press MENU button to enter the horizontal setting system.
Figure 2-31 Table 2-16 The Horizontal menu
Menu
Settings
Comments
Delayed
ON
OFF
Turn on Delayed Scan mode.
Turn off the Delayed Scan mode.
Time Base
Y-T
X-Y
Roll
[1]
Y-axis indicates vertical voltage,
X-axis indicates horizontal time.
Y-axis indicates voltage of CH2,
X-axis indicates voltage of CH1.
The waveform display updates
from right to left.
Trig-offset Reset Reset
Adjust to the center.
NOTE:
[1] No trigger or horizontal offset control of waveforms is available
during Roll Mode, and it‟s only available when set to 500ms/div
or slower.
Term Explanations:
Slow Scan Mode:
This mode is available when the horizontal time base is
set to 50ms or slower. In this mode, the oscilloscope acquires sufficient
data for the left part to the trigger point, then wait for trigger, when trigger
occurs, it continue to draw the rest part from the trigger point to the end of
the right side. When choosing this mode to view low frequency signals, it is
recommended that the channel coupling should be set as DC.