RIGOL
User‟s Guide for DS1000CA Series
2-78
NOTE:
The results of the automatic measurements will be displayed on the bottom
of the screen. Maximum 3 results could be displayed at the same time. When there
is no room, the next new measurement result will make the previous results moving
left out of screen.
Operation Explanation:
1.
Select the signal channel for measuring. CH1 or CH2 according to the signal of
interest.
Press soft buttons as follows: Measure
→
Source
→
CH1 or CH2.
2.
To see all measurement values, set the Display All to ON. 18 measurement
parameters will be displayed on the screen (except for “Delay1
→
2 ” and
“
Delay1
→
2 ” ).
3.
Select parameters page for measuring; select voltage or time parameters pages
by pressing soft button as follows: Measure
→
Voltage or time
→
Vmax, Vmin……
4.
To get the measured value on the screen; select the parameters of interest by
pressing the soft button on the right of the menu, and read the data on the
bottom of the screen.
If the data is displayed as “*****”, it means the parameter cannot be measured
in current condition.
5.
Clear the measure values: press Clear and all of the auto measure values would
disappear from the screen (except for “Display all” parameters).
Figure 2-111 Measure menu
Turn on/off the mass measurement
Clear measured data from screen