Automatic measurement of time parameters
DS1000CA series oscilloscopes provide timing parameters auto-measurements;
Frequency, Period, Rise Time, Fall Time, +Width, -Width, Delay 1
→
2 , Delay 1
→
2 ,
+Duty and -Duty.
Figure 2-113 shows a pulse with some of the time measurement points.
Figure 2-113
The time parameters
Rise Time:
Time that the leading edge of the first pulse in the waveform takes to
rise from 10% to 90% of its amplitude.
Fall Time:
Time that the falling edge of the first pulse in the waveform takes to fall
from 90% to 10% of its amplitude.
+Width:
The width of the first positive pulse in 50% amplitude points.
-Width:
The width of the first negative pulse in the 50% amplitude points.
Delay 1
→
2 :
The delay between the two channels at the rising edge.
Delay 1
→
2 :
The delay between the two channels at the falling edge.
+Duty:
+Duty Cycle, defined as +Width/Period
-Duty:
-Duty Cycle, defined as -Width/Period
10%
50%
90%
-Width
+Width
Fall Time