RIGOL
User‟s Guide for DS1000CA Series
2-79
Automatic measurement of voltage parameters
The DS1000CA series oscilloscopes provide automatic voltage measurements
including Vpp, Vmax, Vmin, Vavg, Vamp, Vrms, Vtop, Vbase, Overshoot and
Preshoot. Figure 2-112 below shows a pulse with some of the voltage measurement
points.
Figure 2-112
Voltage parameters
Vpp:
Peak-to-Peak Voltage.
Vmax:
The maximum amplitude. The most positive peak voltage measured over
the entire waveform.
Vmin:
The minimum amplitude. The most negative peak voltage measured over the
entire waveform.
Vamp:
Voltage between Vtop and Vbase of a waveform
Vtop:
Voltage of the waveform‟s flat top, useful for square/pulse waveforms.
Vbase:
Voltage of the waveform‟s flat base, useful for square/pulse waveforms.
Overshoot:
Defined as (Vmax-Vtop)/Vamp, useful for square and pulse
waveforms.
Preshoot:
Defined as (Vmin-Vbase)/Vamp, useful for square and pulse waveforms.
Average:
The arithmetic mean over the entire waveform.
Vrms:
The true Root Mean Square voltage over the entire waveform.
Vtop
Vmin
Vmax
Overshoot
Vamp
Vbase
Vpp
Preshoot